Diagnostic Tests 6-3
Selection
Description
Default Diags (all)
Runs all the tests.
Tests
Presents a menu of tests to be selected.
Test Groups
Presents a menu of test groups to be selected.
Over Night Loop
Runs selected tests continuously. Only stops when
an error occurs or when the system is reset.
Print Toggle
Turn printing of test groups/tests on or off.
Single Step Toggle
Turns single-stepping of test groups/test on or off.
Looping Toggle
Turns looping of selected test groups/tests on or off.
Error Report Toggle
Turns printing of error statements on or off.
Report Error Log
Prints the first 11 errors that occurred.
Clear Error Log
Erases the buffer of errors.
By default, single-stepping, looping, and test message printing
are turned off while error reporting is turned on. The selected
diagnostics run uninterrupted until they complete or until an error
occurs. If an error has occurred, an error message is printed to
the screen. The message displays the test number, group
number, value expected, and value received. Please contact
Racal-Dana for further interpretation of diagnostic error
messages.
Suppress the error reporting with the
e
command. With error
reporting turned off, selected tests run to completion without
being interrupted by error messages. The 1260-00C indicates if
any errors have occurred after all tests selected have completed.
To view the errors, select
r
to display the error log and
c
to clear
the error log.
Single Step Toggle is used to pinpoint problems. Select
s
to
toggle this command on or off. With this feature, each access to
memory or to a register is reported on the screen. The
1260-00C waits for a key to be pressed before performing the
displayed step.
Diagnostic Test
Selection
If the Default Diags (all) option or the Over Night Loop option is
selected, the appropriate tests begin immediately. The default
option performs all of the tests, while the Over Night Loop option
performs all except the interactive tests and tests that drive
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com