![Quickie R-net Instructions For Use Manual Download Page 52](http://html1.mh-extra.com/html/quickie/r-net/r-net_instructions-for-use-manual_773034052.webp)
Fig.15.
28
Fig.15.
29
Fig.15.
30
15.6.
3.1
Bluetooth Screen Programming
The text to describe each device can be set via the
programmable parameter, Device Name.
The screen symbol for each device can be set via the
programmable parameter, Screen Graphic.
Refer to your healthcare professional or authorized dealer.
15.6.3.2
Disconnecting a Bluetooth Device
Enter the Settings Menu and select Bluetooth. Set the device
you wish to disconnect from On to Off.
15.6.3.3
Flight-Safe Mode
It is common practice to disable wireless transmissions while
on an aircraft. Enter the Settings Menu and select Bluetooth.
Set all the devices to Off.
15.6.4 General Information Symbols
15.6.4.1 Limp (Fig 15.28)
Enter the Settings Menu and select Bluetooth. Set the device
you wish to disconnect from On to Off.
15.6.4.2 Timer (Fig 15.29)
This symbol is displayed when the Omni2 is changing
between different states. An example would be entering into
module re-configuration.
15.6.
4.3
Restart (Fig 15.30)
When the Omni2 requires a reboot; for example, after a
module re-configuration, this symbol will be flashed.
15.6.4.4
Sleep (Fig 15.31)
This symbol will be displayed for a short time before the
Omni2 enters into a sleep state.
15.6.4.5 Cross and Check (Fig 15.32)
T
hese symbols will be displayed during configuration
procedures.
15.6.4.6 E-Stop (Fig 15.33)
If the user switch is activated during drive, or actuator
operation, this symbol will be displayed.
15.6.4.7 SID Displaced (Fig 15.34)
If the SID is operated before or just after you switch the
control system on, the screen will flash the joystick displaced
screen.
You must release and center (where applicable) the SID to
resume normal operation. If you do not release the SID within
five seconds the wheelchair will not be able to move, even if
you release the SID and operate it again. The screen will
display a diagnostic screen at this time. You can reset this
condition by switching the Omni2 off and on again.
15.6.4.8 Locked (Fig 15.35)
This symbol is displayed if the Omni2 is locked.
To unlock the Omni2, turn the power on. When the padlock
symbol appears, press and hold the top navigation button till
you hear a beep and then press and hold the bottom
navigation button till you hear a beep and then it is unlocked.
Fig.15.
31
Fig.15.
33
Process completed
correctly.
Process not completed
correctly.
Fig.15.
32
Fig.15.
34
Fig.15.
35
249044 Rev. A
52
R-Net Control System w/Omni2
Summary of Contents for R-net
Page 1: ...Instructions for Use R Net Control System R Net Controls P N 249044 Rev A ...
Page 7: ...LED and CJSM1 249044 Rev A 7 R Net Control System w Omni2 ...
Page 34: ...Omni2 249044 Rev A 34 R Net Control System w Omni2 ...
Page 76: ...CJSM2 249044 Rev A 76 R Net Control System w Omni2 ...
Page 105: ...249044 Rev A 105 R Net Control System w Omni2 ...
Page 107: ...249044 Rev A 107 R Net Control System w Omni2 ...