Scalar i6000 Installation Guide Super Doc
39
6
Connect the W2 Ethernet cable to the J3 connection on the LBX board on the control module or last
expansion module in the existing configuration.
This connects the J4 connection on the LBX board in the new expansion module to the J3 connection
on the LBX board in the control module or last expansion module in the existing configuration.
7
Install the LBX terminator into the J17 connection on the LBX board in the last expansion module.
8
Use the cover plate thumbscrew to replace the LBX/IEX cover plates on the modules.
LBX board on
expansion module
LBX board on
control module or
W1 cable
J2
J1
J2
J17
J1
expansion module
(large ribbon cable)
J3
J4
W2 cable (Ethernet)
J17 connection
LBX terminator
LBX/IEX cover
plate thumbscrew