Alarm Recording
6-14 Recording Functions
Re
c
o
rd
ing
Func
tions
Alarm
Recording
Priorities
In a delayed recording, waves are recorded in the order assigned in the
Delayed/Alarm Recording Task Window, and the number of waves
recorded is determined by the recorder configured for delayed
recordings. In alarm recordings, the alarming parameters replace the
waves in the lower channels. In the default alarm recording, the top
channel contains the first delayed wave, usually ECG. Additional
delayed channels are overwritten by the additional alarming parameter,
starting at the bottom channel, in the following priority:
•
Parameter whose alarm status has escalated in the last second,
and has alarm recording enabled.
•
Parameter in red alarm with alarm recording enabled.
•
Parameter in yellow alarm with alarm recording enabled.
•
Parameter in red alarm with alarm recording disabled (if space is
available).
•
Parameter in yellow alarm with alarm recording disabled (if space
is available).
Alarm
Recording
Examples
On the Plug-In (2-channel) recorder with respiration recordings
disabled, an apnea alarm followed by a Pressure 1 limit alarm would
produce a strip:
provided that P1 has been selected in the Delayed/Alarm Recordings
Task Window.
For a 4-channel bedside recorder with a standard delayed recording
configured as:
Channel 1
ECG
Channel 2
Pressure 1
Channel 1
ECG
Channel 2
Pressure 1
Channel 3
Pressure 2
Summary of Contents for M1165
Page 12: ...Responsibility of the Manufacturer xii ...
Page 62: ...Using an ITE Display 1 40 The CMS and V24 and V26 Patient Monitors ...
Page 74: ...Attaching the Patient 2 12 Getting Started Getting Started ...
Page 172: ...Alarm Setup 5 16 Alarm Functions Alarm Functions ...
Page 228: ...Loading Paper 6 56 Recording Functions ...
Page 236: ...Admitting a Patient 7 8 Admit Discharge End Case Admit Discharge End Case OR Mode ...
Page 238: ...Admitting a Patient 7 10 Admit Discharge End Case Admit Discharge End Case endcase tif ...
Page 274: ...Drug Calculator 8 36 Trends and Calculations Trends and Calculations ...
Page 299: ...Data Transfer Module Data Transfer 10 3 Data Transfer M1235A CTS DTM CMS CMS V24 V26 ...
Page 388: ...Performance Assurance Checks 13 22 Maintenance Maintenance ...