oxyCRG Display
3-26 Setting up your Monitor
S
e
tt
ing up y
our
M
onito
r
Approximately the bottom two thirds of the Main Screen area are
covered with the three oxyCRG waveforms.
The number of realtime waves that you can display simultaneously with
oxyCRG is related to the maximum number of realtime waves that your
system is capable of displaying. Details are provided in the following
table:
Main Screen
Notes on
oxyCRG
•
The oxyCRG information is overlayed on the realtime waves at the
bottom of the Main Screen. These realtime waves disappear from
the display, but the parameters are still being monitored, and the
parameter alarms are still active.
•
If oxyCRG covers part of an overlap channel, the entire overlap
channel disappears from the display.
•
The
softkey has no effect on the oxyCRG section
of the Main Screen, because the display speed is very low.
•
The beat-to-beat heart rate (btbHR) in the oxyCRG display will
vary from the heart rate (HR) numeric on the Main Screen, since
the HR numeric is an averaged value.
•
If the monitor is switched off for less than 60 seconds, oxyCRG
data is retained. However, when the power is switched back on,
you will see gaps in the oxyCRG display. The gaps correspond to
the amount of time it takes the monitor to get the first readings
after power is restored,
not
the amount of time that the monitor
was switched off. For this reason, you cannot properly measure
the time difference between an event which occurs before power
Standard Display
oxyCRG Display
4 realtime waves
1 realtime wave
6 realtime waves
2 realtime waves
8 realtime waves
3 realtime waves
Freeze Waves
Summary of Contents for M1165
Page 12: ...Responsibility of the Manufacturer xii ...
Page 62: ...Using an ITE Display 1 40 The CMS and V24 and V26 Patient Monitors ...
Page 74: ...Attaching the Patient 2 12 Getting Started Getting Started ...
Page 172: ...Alarm Setup 5 16 Alarm Functions Alarm Functions ...
Page 228: ...Loading Paper 6 56 Recording Functions ...
Page 236: ...Admitting a Patient 7 8 Admit Discharge End Case Admit Discharge End Case OR Mode ...
Page 238: ...Admitting a Patient 7 10 Admit Discharge End Case Admit Discharge End Case endcase tif ...
Page 274: ...Drug Calculator 8 36 Trends and Calculations Trends and Calculations ...
Page 299: ...Data Transfer Module Data Transfer 10 3 Data Transfer M1235A CTS DTM CMS CMS V24 V26 ...
Page 388: ...Performance Assurance Checks 13 22 Maintenance Maintenance ...