Parameter Settings Transfer Messages
Setting up your Monitor 3-63
S
e
tt
ing up y
our
M
onit
o
r
Parameter Settings Transfer Messages
The following messages related to Parameter Settings Transfer can
appear on the display:
Message
Condition
Action required
(Parameter name)
no settings
transfer capability
The indicated parameter
module is not a
T
module.
Insert a module that has
a
T
on the front.
(Parameter name)
check settings''
No settings are stored in
the module, or the
settings are incorrectly
stored.
Enter the parameter
Task Window and check
the settings. The
settings are
automatically stored
when you exit the Task
Window.
Check Patient
Category and
(parameter name)
settings
The patient category
stored with the settings
in the module is
different from that of the
monitor.
Either
change the
patient category for the
monitor
or
change the
parameter settings.
(Parameter name)
unable to store
settings in module
The parameter module
was removed before the
settings could be stored,
or there is a malfunction
in the module. (This
message can appear with
Settings Transfer On or
Off.)
Press ,
re-insert the module and
check/adjust the
settings in the
parameter Task
Window. Exit the Task
Window and leave the
module in the rack for
about 10 seconds before
you remove it. If the
same message appears,
replace the module.
Silence Reset
Summary of Contents for M1165
Page 12: ...Responsibility of the Manufacturer xii ...
Page 62: ...Using an ITE Display 1 40 The CMS and V24 and V26 Patient Monitors ...
Page 74: ...Attaching the Patient 2 12 Getting Started Getting Started ...
Page 172: ...Alarm Setup 5 16 Alarm Functions Alarm Functions ...
Page 228: ...Loading Paper 6 56 Recording Functions ...
Page 236: ...Admitting a Patient 7 8 Admit Discharge End Case Admit Discharge End Case OR Mode ...
Page 238: ...Admitting a Patient 7 10 Admit Discharge End Case Admit Discharge End Case endcase tif ...
Page 274: ...Drug Calculator 8 36 Trends and Calculations Trends and Calculations ...
Page 299: ...Data Transfer Module Data Transfer 10 3 Data Transfer M1235A CTS DTM CMS CMS V24 V26 ...
Page 388: ...Performance Assurance Checks 13 22 Maintenance Maintenance ...