Procedure Recordings
6-18 Recording Functions
Re
c
o
rd
ing
Func
tions
→
→
See the Pressure Section for details on Pulmonary Artery Wedge
Pressure (PAWP) Measurements.
When a PAWP measurement has been performed, the PAWP waveform
can be recorded by pressing the
softkey. If displayed,
the CO
2
or RESP waveform is also recorded.
The recording strip is also annotated with the date, time, and wedge
value.
ST
Recordings
Recordings of ST waves and their reference waves can be made on the
bedside or central recorders. They are initiated from the ST Analysis
Task Window. An ST recording shows the ST beat(s) from the cursor
spot in the upper channel, and the corresponding reference beat(s) in
the lower channel, with the measurement values, net change, and
measurement points listed in the annotation.
ST values transferred by the Philips M1235A Data Transfer Module are
indicated by a “T” after the time. Because the ST waveforms are not
transferred, no waveforms will be recorded.
Procedures
Wedge
Edit Wedge
Record Curve
Summary of Contents for M1165
Page 12: ...Responsibility of the Manufacturer xii ...
Page 62: ...Using an ITE Display 1 40 The CMS and V24 and V26 Patient Monitors ...
Page 74: ...Attaching the Patient 2 12 Getting Started Getting Started ...
Page 172: ...Alarm Setup 5 16 Alarm Functions Alarm Functions ...
Page 228: ...Loading Paper 6 56 Recording Functions ...
Page 236: ...Admitting a Patient 7 8 Admit Discharge End Case Admit Discharge End Case OR Mode ...
Page 238: ...Admitting a Patient 7 10 Admit Discharge End Case Admit Discharge End Case endcase tif ...
Page 274: ...Drug Calculator 8 36 Trends and Calculations Trends and Calculations ...
Page 299: ...Data Transfer Module Data Transfer 10 3 Data Transfer M1235A CTS DTM CMS CMS V24 V26 ...
Page 388: ...Performance Assurance Checks 13 22 Maintenance Maintenance ...