The Event Setup task window enables you to adjust the following items:
Event
Criteria
The event type (Apnea, Bradycardia, Desaturation) determines how a
specific event is triggered and stored on the monitor.
Apnea Event
Settings
Apnea events are
always
triggered by a Red*** Apnea alarm. The Apnea
alarm is derived from the RESP parameter and occurs whenever
breathing ceases for a period exceeding the defined Apnea alarm limit
(usually 20 seconds).
The Apnea alarm limit and trigger time can be viewed but not adjusted in
the Event Setup task window. These settings must be adjusted in
Monitoring Mode
, in the RESP task window.
Trigger Mode
The trigger mode determines whether an
event is triggered by a physiological alarm or
by criteria defined specifically by the user.
Trig Threshold
The trigger threshold can be adjusted if the
user defined trigger mode setting has been
selected
Trigger Time
The trigger time can be adjusted if the user
defined trigger mode setting has been
selected.
Summary of Contents for M1165
Page 12: ...Responsibility of the Manufacturer xii ...
Page 62: ...Using an ITE Display 1 40 The CMS and V24 and V26 Patient Monitors ...
Page 74: ...Attaching the Patient 2 12 Getting Started Getting Started ...
Page 172: ...Alarm Setup 5 16 Alarm Functions Alarm Functions ...
Page 228: ...Loading Paper 6 56 Recording Functions ...
Page 236: ...Admitting a Patient 7 8 Admit Discharge End Case Admit Discharge End Case OR Mode ...
Page 238: ...Admitting a Patient 7 10 Admit Discharge End Case Admit Discharge End Case endcase tif ...
Page 274: ...Drug Calculator 8 36 Trends and Calculations Trends and Calculations ...
Page 299: ...Data Transfer Module Data Transfer 10 3 Data Transfer M1235A CTS DTM CMS CMS V24 V26 ...
Page 388: ...Performance Assurance Checks 13 22 Maintenance Maintenance ...