Alarm Setup
Alarm Functions 5-15
Ala
rm Func
tions
The Nurse
Call Relay
Note—
The nurse call relay is not available for the V24CT and V26CT.
The nurse call relay is a paging system which warns the nurse if an alarm
occurs on a monitor. It becomes active in one of four situations
depending on the way it has been configured:
•
Red only alarms
•
Red and Yellow alarms
•
Red and INOP alarms
•
Red and Yellow INOP alarms.
The nurse call relay is not active when the monitor’s alarm capability is
switched off.
Note—
The delay time to the nurse call relay output is max. 1 sec.
When an alarm has been silenced using the
key, and the
audible 2
or
3 minute alarm reminder is configured to be either On or Re-
alarm, the nurse call relay signals will behave as follows:
Alarm
Reminder
Reminder
Time
Nurse Call Relay Signal Behavior
On
2
or
3 mins
The nurse call relay signals are repeated
for a duration of 6 seconds every 2 or 3
minutes unless the alarm condition is
eliminated
Re-alarm
2
or
3 mins
The nurse call relay signals are activated
after 2 or 3 minutes. The alarm tone and
relay signals run continuously until they
are interrupted by pressing the
key.
Silence/Reset
Silence/Reset
Summary of Contents for M1165
Page 12: ...Responsibility of the Manufacturer xii ...
Page 62: ...Using an ITE Display 1 40 The CMS and V24 and V26 Patient Monitors ...
Page 74: ...Attaching the Patient 2 12 Getting Started Getting Started ...
Page 172: ...Alarm Setup 5 16 Alarm Functions Alarm Functions ...
Page 228: ...Loading Paper 6 56 Recording Functions ...
Page 236: ...Admitting a Patient 7 8 Admit Discharge End Case Admit Discharge End Case OR Mode ...
Page 238: ...Admitting a Patient 7 10 Admit Discharge End Case Admit Discharge End Case endcase tif ...
Page 274: ...Drug Calculator 8 36 Trends and Calculations Trends and Calculations ...
Page 299: ...Data Transfer Module Data Transfer 10 3 Data Transfer M1235A CTS DTM CMS CMS V24 V26 ...
Page 388: ...Performance Assurance Checks 13 22 Maintenance Maintenance ...