Performing and Reviewing Calculations
8-24 Trends and Calculations
Tr
e
nds
a
nd
Calcula
tion
s
Resampling
Vital SIgns
Resampling of vital signs enables you to override the values in the
Trends database, which are stored every 12 seconds or averaged over 1
minute, depending on the monitor’s configuration. Resampling enables
you to use the most recent continuously monitored values. Resampling
sets the calculation time to the current time, and displays the
corresponding values for the previous second in the calculations task
window. Resampled values are indicated by a preceding asterisk (*) in
Calculations, Graph Trends, and Vital Signs task windows.
If no current values exist, then the most recent value within 30 minutes
will be retrieved. The most recent height, weight, (and head
circumference for neonatal only) are used regardless of how long ago
they were entered.
Pressing
stores the resampled values at the current time.
Pressing
also stores the resampled values at the current
time and does the calculations using the resampled values. Pressing any
other softkey except
cancels the resampled values and
brings up the database values for the displayed calculation time.
Resampled values that are not stored or used in calculations will be
erased.
When data is transferred with the Data Transfer module, values that
have been resampled are transferred to the destination monitor.
Resampled and other manually entered data is annotated with an
asterisk (*) at the destination monitor if both the source and destination
monitors have software at the Release F level or higher (CMS) or
Release A.0 level or higher (V24 and V26).
Confirm
Perform Calc
On/Off DataTime
Summary of Contents for M1165
Page 12: ...Responsibility of the Manufacturer xii ...
Page 62: ...Using an ITE Display 1 40 The CMS and V24 and V26 Patient Monitors ...
Page 74: ...Attaching the Patient 2 12 Getting Started Getting Started ...
Page 172: ...Alarm Setup 5 16 Alarm Functions Alarm Functions ...
Page 228: ...Loading Paper 6 56 Recording Functions ...
Page 236: ...Admitting a Patient 7 8 Admit Discharge End Case Admit Discharge End Case OR Mode ...
Page 238: ...Admitting a Patient 7 10 Admit Discharge End Case Admit Discharge End Case endcase tif ...
Page 274: ...Drug Calculator 8 36 Trends and Calculations Trends and Calculations ...
Page 299: ...Data Transfer Module Data Transfer 10 3 Data Transfer M1235A CTS DTM CMS CMS V24 V26 ...
Page 388: ...Performance Assurance Checks 13 22 Maintenance Maintenance ...