Recording Status Messages
Recording Functions 6-47
Re
c
o
rd
ing
Func
tions
Note—
Recording ink sometimes fades when covered with transparent
tape. Please avoid covering any part of a recording that is clinically
relevant (annotation or waveforms) when taping a recording strip to a
patient record or other patient documentation.
Wrong Procedure
Recorder Configuration
A trended vital signs recording or neonatal
event recording has been requested but not
supported by the selected procedure
recorder.
Check Paper/Door on
Recorder Name
Recorder|
The bedside recorder is out of paper or the
recorder door is open - load new paper and/
or shut the door.
Recorder Name
Recorder
Stopped
An actively running recording has been
stopped by pressing
on the recorder
or the
softkey on the
display or a timed recording has stopped
automatically.
Recording:
Waves that
are being Rec.
A timed recording is currently running on a
Bedside or a Central Recorder.
Waiting for
Recorder
Name
Recorder
A recording has been requested for a
recorder that is currently busy.
Recorder does not
support
Recording Type
The recorder configured cannot support the
requested application. For example, an
oxyCRG recording that is requested on a
recorder other than M1116B.
Running Recordings
Mode
A list of running recordings is shown when
is pressed.
Message
Meaning
STOP
Stop Recordng
Record
Summary of Contents for M1165
Page 12: ...Responsibility of the Manufacturer xii ...
Page 62: ...Using an ITE Display 1 40 The CMS and V24 and V26 Patient Monitors ...
Page 74: ...Attaching the Patient 2 12 Getting Started Getting Started ...
Page 172: ...Alarm Setup 5 16 Alarm Functions Alarm Functions ...
Page 228: ...Loading Paper 6 56 Recording Functions ...
Page 236: ...Admitting a Patient 7 8 Admit Discharge End Case Admit Discharge End Case OR Mode ...
Page 238: ...Admitting a Patient 7 10 Admit Discharge End Case Admit Discharge End Case endcase tif ...
Page 274: ...Drug Calculator 8 36 Trends and Calculations Trends and Calculations ...
Page 299: ...Data Transfer Module Data Transfer 10 3 Data Transfer M1235A CTS DTM CMS CMS V24 V26 ...
Page 388: ...Performance Assurance Checks 13 22 Maintenance Maintenance ...