Maintenance Checks
11-14 Monitor Installation and Patient Safety
M
onitor
Ins
ta
lla
ti
on
a
nd Pa
ti
e
n
t Sa
fe
ty
Maintenance Checks
Before commencing monitoring on a patient, carry out the following
checks on the system:
•
Check for any mechanical damage
•
Check all the external leads, plug-ins and accessories
•
Check all the functions of the instrument which will be needed to
monitor the patient, and ensure that the instrument is in good
working order.
Do not use the system for any monitoring procedure on a patient
if you identify features which demonstrate impaired functioning of the
instrument. Contact the hospital biomedical engineer or the Philips
Service Engineer.
Maintenance
Frequency
Source of Information
Full performance checks
including safety checks.
At least every two years, after
any repairs, or as needed.
See “Testing and Maintaining the
CMS” in the “CMS Service Guide”
or “Testing and Maintaining the
V24 and V26“in the “V24 and V26
Service Guide”.
Synchronization of the monitor
and defibrillator.
If defibrillator is used or
configured for intended use: at
least once a year, after any
repairs, or as needed.
See the Philips M1722/3/4A
Codemaster XL/XL+/XE
Defibrillator Service Guide.
Preventive maintenance
Perform the NBP and Sidestream
CO 2 Module Performance test
blocks every year.
Perform the Preventive
Maintenance Procedure for the
M1180A #A01/A02 Wall Mount
Assembly every year.
See “Testing and Maintaining the
CMS” in the “CMS Service Guide”
or “Testing and Maintaining the
V24 and V26“in the “V24 and V26
Service Guide”.
Summary of Contents for M1165
Page 12: ...Responsibility of the Manufacturer xii ...
Page 62: ...Using an ITE Display 1 40 The CMS and V24 and V26 Patient Monitors ...
Page 74: ...Attaching the Patient 2 12 Getting Started Getting Started ...
Page 172: ...Alarm Setup 5 16 Alarm Functions Alarm Functions ...
Page 228: ...Loading Paper 6 56 Recording Functions ...
Page 236: ...Admitting a Patient 7 8 Admit Discharge End Case Admit Discharge End Case OR Mode ...
Page 238: ...Admitting a Patient 7 10 Admit Discharge End Case Admit Discharge End Case endcase tif ...
Page 274: ...Drug Calculator 8 36 Trends and Calculations Trends and Calculations ...
Page 299: ...Data Transfer Module Data Transfer 10 3 Data Transfer M1235A CTS DTM CMS CMS V24 V26 ...
Page 388: ...Performance Assurance Checks 13 22 Maintenance Maintenance ...