PG D
RIVES
T
ECHNOLOGY
O
MNI
– D
IAGNOSTICS
SK78813/5
153
• Check all cables and connectors between the Omni and the User Switch.
If the trip is still present after the above checks have been made, then the User
Switch may be defective. Contact your service agent.
If it is required to use the Omni without a User Switch being connected, then the
parameter, Switch Detect, should be set to Off. If a User Switch is not used the
responsibility for that decision lies with the healthcare professional.
Because a disconnected User Switch means there is
no emergency stop function, PG Drives Technology
recommend that Switch Detect is this option is
always set to On. PG Drives Technology will accept
no liability for any losses resulting from any other
setting of this parameter.
3.25 Over Pressure
The Omni has detected that the pressure on the Sip and Puff transducer has
exceeded its maximum level. Although a trip screen is not displayed the system
log will show the trip and number of occurrences.
•
Check the user is operating the Sip and Puff device correctly.
If the trip is still present after the above check has been made, then the Omni
may be defective. Refer to section 5.
3.26 Gone to Sleep
This occurs when the control system has been left inactive for a time greater
than the parameter Sleep Timer.
An entry is made in the system log each time this occurs.
3.27 Charging
This occurs when the control system detects that a charger is connected to
either Inhibit 1 or Inhibit 3. Refer to section 2.3 for connection details
The Battery charging screen will be displayed during charger connection.
An entry is made in the system log each time this occurs.
Summary of Contents for R-NET OMNI
Page 1: ...PG DRIVES TECHNOLOGY R NET OMNI SK78813 5 1 R NET OMNI TECHNICAL MANUAL SK78813 5 ...
Page 16: ...PG DRIVES TECHNOLOGY R NET OMNI SK78813 5 16 ...
Page 17: ...PG DRIVES TECHNOLOGY OMNI STANDARD INSTALLATION SK78813 5 17 CHAPTER 1 INSTALLATION ...
Page 18: ...PG DRIVES TECHNOLOGY OMNI STANDARD INSTALLATION SK78813 5 18 ...
Page 20: ...PG DRIVES TECHNOLOGY OMNI STANDARD INSTALLATION SK78813 5 20 2 Controls and Connections ...
Page 39: ...PG DRIVES TECHNOLOGY OMNI STANDARD OPERATION SK78813 5 39 CHAPTER 2 OPERATION ...
Page 40: ...PG DRIVES TECHNOLOGY OMNI STANDARD OPERATION SK78813 5 40 ...
Page 68: ...PG DRIVES TECHNOLOGY OMNI STANDARD OPERATION SK78813 5 68 ...
Page 69: ...PG DRIVES TECHNOLOGY OMNI STANDARD PROGRAMMING SK78813 5 69 CHAPTER 3 PROGRAMMING STANDARD ...
Page 70: ...PG DRIVES TECHNOLOGY OMNI STANDARD PROGRAMMING SK78813 5 70 ...
Page 76: ...PG DRIVES TECHNOLOGY OMNI STANDARD PROGRAMMING SK78813 5 76 1 3 1 PC Programmer Parameters ...
Page 99: ...PG DRIVES TECHNOLOGY OMNI ADVANCED PROGRAMMING SK78813 5 99 CHAPTER 4 PROGRAMMING ADVANCED ...
Page 100: ...PG DRIVES TECHNOLOGY OMNI ADVANCED PROGRAMMING SK78813 5 100 ...
Page 109: ...PG DRIVES TECHNOLOGY OMNI IR SK78813 5 109 CHAPTER 5 INFA RED CONTROL ...
Page 110: ...PG DRIVES TECHNOLOGY OMNI IR SK78813 5 110 ...
Page 135: ...PG DRIVES TECHNOLOGY OMNI DIAGNOSTICS SK78813 5 135 CHAPTER 5 DIAGNOSTICS ...
Page 136: ...PG DRIVES TECHNOLOGY OMNI DIAGNOSTICS SK78813 5 136 ...
Page 141: ...PG DRIVES TECHNOLOGY OMNI DIAGNOSTICS SK78813 5 141 ...
Page 158: ...PG DRIVES TECHNOLOGY OMNI DIAGNOSTICS SK78813 5 158 ...
Page 159: ...PG DRIVES TECHNOLOGY OMNI SPECIFICATIONS SK78813 5 159 CHAPTER 6 SPECIFICATIONS ...
Page 160: ...PG DRIVES TECHNOLOGY OMNI SPECIFICATIONS SK78813 5 160 ...
Page 167: ...PG DRIVES TECHNOLOGY OMNI WARNING SUMMARY SK78813 5 167 CHAPTER 7 WARNING SUMMARY ...
Page 168: ...PG DRIVES TECHNOLOGY OMNI WARNING SUMMARY SK78813 5 168 ...