Troubleshooting
285
•
Be sure that the injector you are using is recommended for the types of solvents
you are using. The quartz injector, 1.2 mm i.d., is recommended for typical
organic analyses. The alumina injector, 0.80 mm i.d., should be used with
volatile organic solvents, for example, methanol.
•
Try increasing the RF power, using less solvent, slowing down the pump speed,
or adjusting the torch height.
Printing Problems
Table 6-2.
Printing Problems
Problem
Possible Cause(s)
Corrective Action
Printer does not print
The correct printer driver may
not be installed or selected in
Windows.
Check the printer drivers in Windows.
For more information, see the Syngistix
Software installation Guide on the
Syngistix Document Pack CD.
The printer may have been
disconnected.
Check the power cable and the cable
connecting the printer to the computer.
Make sure the power is on.
The printer may be out of
paper or off-line.
Check printer status.
The pins on the printer cable
connector may be broken.
Check the connector. If the pins are
broken, order a replacement cable.
Printout is missing
information
The printer may not be set to
the correct emulation mode.
See the printer manufacturer’s
information or refer to the
Syngistix
Software Installtion Guide on the
Syngistix Document Pack CD
.
Summary of Contents for AVIO 200
Page 1: ...AVIO 200 SPECTROMETER Hardware Guide ICP OPTICAL EMISSION ...
Page 2: ......
Page 3: ...AvioTM 200 Spectrometer Customer Hardware and Service Guide ...
Page 12: ...Contents 10 ...
Page 30: ...28 ...
Page 31: ...Safety Practices 1 ...
Page 32: ...30 Safety Practices ...
Page 56: ...54 Safety Practices ...
Page 57: ...Preparing Your Laboratory 2 ...
Page 58: ...Preparing Your Laboratory 56 ...
Page 70: ...Preparing Your Laboratory 68 ...
Page 71: ...System Description 3 ...
Page 72: ......
Page 97: ...Installation 4 ...
Page 98: ......
Page 119: ...Installation 117 Figure 4 8 Replacing the torch 1 3 4 6 8 2 9 10 5 7 ...
Page 121: ...Installation 119 1 2 7 9 8 10 3 4 5 6 ...
Page 164: ...Installation 162 ...
Page 165: ...Maintenance 5 ...
Page 166: ......
Page 184: ...Maintenance 182 1 2 7 9 8 10 3 4 5 6 ...
Page 188: ...Maintenance 186 Figure 5 8 Replacing the Torch 1 3 4 6 8 2 9 10 5 7 ...
Page 272: ...Maintenance 270 ...
Page 273: ...Troubleshooting 6 ...
Page 274: ......
Page 293: ...Troubleshooting 291 Figure 6 1 Normal Plasma Conditions at 0 torch position ...
Page 294: ...Troubleshooting 292 Figure 6 2 Normal Plasma at 3mm position ...
Page 297: ...Troubleshooting 295 Figure 6 5 Injector too far forward ...
Page 298: ...Troubleshooting 296 Figure 6 6 No Aux Gas Torch may be glowing ...
Page 299: ...Troubleshooting 297 Figure 6 7 Air Leak or Spray Chamber Temperature too high ...
Page 300: ...Troubleshooting 298 Figure 6 8 Thin Plasma due to leak in Plasma Gas Line ...
Page 308: ......