Maintenance
241
Mira Mist Nebulizer
Caution
Attention
When cleaning the concentric glass nebulizer, do
not
use ultrasonic
cleaning. This can adversely affect nebulizer performance.
Do
not
attempt to clean out the concentric nebulizer with any wire
instrument. This can cause irreparable damage to the nebulizer.
Lors du nettoyage du nébuliseur en verre concentriques, ne pas utiliser le
nettoyage par ultrasons. Cela peut nuire à la performance nébuliseur.
Ne pas
essayer de nettoyer le nébuliseur concentrique avec tout instrument
de fil. Cela peut causer des dommages irréparables au nébuliseur.
This device operates on compressed gases.
Appropriate care
must be taken. If in doubt about correct operating procedures, call
an experienced operator. Burgener Mira Mist Nebulizers require
45 - 55 psi to have a 1 liter per minute of Argon gas flow, so the
operating pressures are in the range of 25 - 45 psi, depending on
the torch optimum flow rate.
Mira Mist nebulizers will pulse if the pump cannot deliver
constant sample flow: Change your pump tubing often, or use a
surgeless pump if possible.
!
Warning
Summary of Contents for AVIO 200
Page 1: ...AVIO 200 SPECTROMETER Hardware Guide ICP OPTICAL EMISSION ...
Page 2: ......
Page 3: ...AvioTM 200 Spectrometer Customer Hardware and Service Guide ...
Page 12: ...Contents 10 ...
Page 30: ...28 ...
Page 31: ...Safety Practices 1 ...
Page 32: ...30 Safety Practices ...
Page 56: ...54 Safety Practices ...
Page 57: ...Preparing Your Laboratory 2 ...
Page 58: ...Preparing Your Laboratory 56 ...
Page 70: ...Preparing Your Laboratory 68 ...
Page 71: ...System Description 3 ...
Page 72: ......
Page 97: ...Installation 4 ...
Page 98: ......
Page 119: ...Installation 117 Figure 4 8 Replacing the torch 1 3 4 6 8 2 9 10 5 7 ...
Page 121: ...Installation 119 1 2 7 9 8 10 3 4 5 6 ...
Page 164: ...Installation 162 ...
Page 165: ...Maintenance 5 ...
Page 166: ......
Page 184: ...Maintenance 182 1 2 7 9 8 10 3 4 5 6 ...
Page 188: ...Maintenance 186 Figure 5 8 Replacing the Torch 1 3 4 6 8 2 9 10 5 7 ...
Page 272: ...Maintenance 270 ...
Page 273: ...Troubleshooting 6 ...
Page 274: ......
Page 293: ...Troubleshooting 291 Figure 6 1 Normal Plasma Conditions at 0 torch position ...
Page 294: ...Troubleshooting 292 Figure 6 2 Normal Plasma at 3mm position ...
Page 297: ...Troubleshooting 295 Figure 6 5 Injector too far forward ...
Page 298: ...Troubleshooting 296 Figure 6 6 No Aux Gas Torch may be glowing ...
Page 299: ...Troubleshooting 297 Figure 6 7 Air Leak or Spray Chamber Temperature too high ...
Page 300: ...Troubleshooting 298 Figure 6 8 Thin Plasma due to leak in Plasma Gas Line ...
Page 308: ......