Maintenance
169
Periodic Checks
The frequency of maintenance required will be determined by the amount of use the
system receives, the cleanliness of the environment, and the number and nature of
the samples being analyzed. You should consider the usage of your spectrometer and
establish a suitable maintenance schedule, with particular emphasis on the following
areas of the system.
Torch Assembly and Plasma Induction Plates
Clean the torch components periodically according to the instructions given later in
this chapter to remove accumulated deposits. If using organic solvents, check the
injector and torch regularly for carbon buildup. Replace any cracked or worn O-rings
on the torch assembly.
The plasma induction plates must be kept clean to prevent arcing across the plasma
induction plates. Inspect the plasma induction plates for any deformations or carbon
buildup. Contact your PerkinElmer Service Representative to replace the plasma
induction plates if there are any signs of pitting. Pitting causes weakness in the
plasma induction plates which can result in a hole in the plasma induction plates and
a gas leak.
Purge Viewing Window/Tubes
Remove and examine the window or purge tube for deposits. If you notice a drop in
UV performance clean or replace the windows. Clean according to the instructions
given later in this chapter or replace as necessary.
Caution
Attention
To extend the life and aid in preventing devitrification of glassware, it is
necessary to remove all traces of alkali on glassware caused by handling it
with bare hands. To remove the alkali, wipe the glassware clean with
cotton wool dipped in alcohol.
Pour prolonger la durée de vie et de l'aide dans la prévention de la
dévitrification de la verrerie, il est nécessaire d'éliminer toute trace d'alcali
sur la verrerie provoquée par la manipulation à mains nues. Pour supprimer
l'alcali, essuyez la verrerie propre avec un coton imbibé d'alcool.
Summary of Contents for AVIO 200
Page 1: ...AVIO 200 SPECTROMETER Hardware Guide ICP OPTICAL EMISSION ...
Page 2: ......
Page 3: ...AvioTM 200 Spectrometer Customer Hardware and Service Guide ...
Page 12: ...Contents 10 ...
Page 30: ...28 ...
Page 31: ...Safety Practices 1 ...
Page 32: ...30 Safety Practices ...
Page 56: ...54 Safety Practices ...
Page 57: ...Preparing Your Laboratory 2 ...
Page 58: ...Preparing Your Laboratory 56 ...
Page 70: ...Preparing Your Laboratory 68 ...
Page 71: ...System Description 3 ...
Page 72: ......
Page 97: ...Installation 4 ...
Page 98: ......
Page 119: ...Installation 117 Figure 4 8 Replacing the torch 1 3 4 6 8 2 9 10 5 7 ...
Page 121: ...Installation 119 1 2 7 9 8 10 3 4 5 6 ...
Page 164: ...Installation 162 ...
Page 165: ...Maintenance 5 ...
Page 166: ......
Page 184: ...Maintenance 182 1 2 7 9 8 10 3 4 5 6 ...
Page 188: ...Maintenance 186 Figure 5 8 Replacing the Torch 1 3 4 6 8 2 9 10 5 7 ...
Page 272: ...Maintenance 270 ...
Page 273: ...Troubleshooting 6 ...
Page 274: ......
Page 293: ...Troubleshooting 291 Figure 6 1 Normal Plasma Conditions at 0 torch position ...
Page 294: ...Troubleshooting 292 Figure 6 2 Normal Plasma at 3mm position ...
Page 297: ...Troubleshooting 295 Figure 6 5 Injector too far forward ...
Page 298: ...Troubleshooting 296 Figure 6 6 No Aux Gas Torch may be glowing ...
Page 299: ...Troubleshooting 297 Figure 6 7 Air Leak or Spray Chamber Temperature too high ...
Page 300: ...Troubleshooting 298 Figure 6 8 Thin Plasma due to leak in Plasma Gas Line ...
Page 308: ......