Troubleshooting
277
Capillary Tubing
Are the nebulizer and sample capillary tubing properly connected? Is the tubing
discolored or clogged? Is tubing crimped or pinched?
If the capillary tubing is not in good condition, replace it.
Nebulizer and Spray Chamber
Is there leakage around the nebulizer? Is there leakage from the spray chamber drain?
Is fluid accumulating in the torch?
Make sure the nebulizer is on securely. If you cannot get a secure fit, the nebulizer
or knurled nut (for Meinhard K1) O-ring may need to be replaced. Check the drain
fitting and tubing to be sure the spray chamber is being properly drained. Check that
the waste is being pumped in the proper direction.
Check the nebulizer spray pattern: run deionized water for several minutes, then
remove the nebulizer. Is the nebulizer spray sputtering or uneven?
Check the nebulizer spray pattern: run deionized water for several minutes, then
remove the nebulizer. If the nebulizer spray is sputtering or uneven, inspect the
nebulizer tips for clogging and clean or replace them as necessary.
Are there deposits inside the end cap or spray chamber?
Clean out deposits. See
Torch Assembly
Are the torch fittings for the gas inlets secure?
If needed tighten the fittings on the torch manifold. Do
not
overtighten them. See
Is the quartz torch cloudy or dirty?
Check for deposits, particularly if running organics, alloys, or samples with high
dissolved solids. See
Is the injector dirty? Is it deformed due to overheating?
The injector may be clogged and require cleaning. If you are running organics, check
for carbon buildup. If running samples with high dissolved solids, check for deposits.
See
If the torch requires disassembly, check the O-rings. Are they cracked or worn? Is
the injector adapter discolored? See
Replace worn parts.
Summary of Contents for AVIO 200
Page 1: ...AVIO 200 SPECTROMETER Hardware Guide ICP OPTICAL EMISSION ...
Page 2: ......
Page 3: ...AvioTM 200 Spectrometer Customer Hardware and Service Guide ...
Page 12: ...Contents 10 ...
Page 30: ...28 ...
Page 31: ...Safety Practices 1 ...
Page 32: ...30 Safety Practices ...
Page 56: ...54 Safety Practices ...
Page 57: ...Preparing Your Laboratory 2 ...
Page 58: ...Preparing Your Laboratory 56 ...
Page 70: ...Preparing Your Laboratory 68 ...
Page 71: ...System Description 3 ...
Page 72: ......
Page 97: ...Installation 4 ...
Page 98: ......
Page 119: ...Installation 117 Figure 4 8 Replacing the torch 1 3 4 6 8 2 9 10 5 7 ...
Page 121: ...Installation 119 1 2 7 9 8 10 3 4 5 6 ...
Page 164: ...Installation 162 ...
Page 165: ...Maintenance 5 ...
Page 166: ......
Page 184: ...Maintenance 182 1 2 7 9 8 10 3 4 5 6 ...
Page 188: ...Maintenance 186 Figure 5 8 Replacing the Torch 1 3 4 6 8 2 9 10 5 7 ...
Page 272: ...Maintenance 270 ...
Page 273: ...Troubleshooting 6 ...
Page 274: ......
Page 293: ...Troubleshooting 291 Figure 6 1 Normal Plasma Conditions at 0 torch position ...
Page 294: ...Troubleshooting 292 Figure 6 2 Normal Plasma at 3mm position ...
Page 297: ...Troubleshooting 295 Figure 6 5 Injector too far forward ...
Page 298: ...Troubleshooting 296 Figure 6 6 No Aux Gas Torch may be glowing ...
Page 299: ...Troubleshooting 297 Figure 6 7 Air Leak or Spray Chamber Temperature too high ...
Page 300: ...Troubleshooting 298 Figure 6 8 Thin Plasma due to leak in Plasma Gas Line ...
Page 308: ......