Maintenance
247
Installing the Sample Tubing
To install the sample tubing refer to the instructions below and the next two figures:
1. Turn off the peristaltic pump using the software control.
2. Take a piece of 0.76 mm (0.030 in.) pump sample tubing. New pump tubing will
work better if it is pre-stretched, and analytical precision will be improved.
Grasp the tubing, with one clip in each hand, and stretch the tubing gently a few
times.
3. Install the pump tubing on an open channel of the pump. To do this, carefully
stretch the new tubing around the pump head and place the clips on the pump
tubing into the slots, using the slots in the same position on each side of the pump
head. Carefully thread the pump tubing through the lip of the cover (see the next
figure). Be careful not to twist or kink the pump tubing during this process.
4. Take two pieces of 0.58 mm i.d. (0.023 in.) polyethylene tubing (Part No.
09908265), one for the sample capillary and one for the nebulizer capillary.
Using a razor blade, bevel one end of each piece of capillary tubing by cutting it
at an angle. See the next figure.
5. Insert the beveled end of the sample capillary tubing into one end of the pump
tubing. Lead the capillary tubing to the sample bottle.
Summary of Contents for AVIO 200
Page 1: ...AVIO 200 SPECTROMETER Hardware Guide ICP OPTICAL EMISSION ...
Page 2: ......
Page 3: ...AvioTM 200 Spectrometer Customer Hardware and Service Guide ...
Page 12: ...Contents 10 ...
Page 30: ...28 ...
Page 31: ...Safety Practices 1 ...
Page 32: ...30 Safety Practices ...
Page 56: ...54 Safety Practices ...
Page 57: ...Preparing Your Laboratory 2 ...
Page 58: ...Preparing Your Laboratory 56 ...
Page 70: ...Preparing Your Laboratory 68 ...
Page 71: ...System Description 3 ...
Page 72: ......
Page 97: ...Installation 4 ...
Page 98: ......
Page 119: ...Installation 117 Figure 4 8 Replacing the torch 1 3 4 6 8 2 9 10 5 7 ...
Page 121: ...Installation 119 1 2 7 9 8 10 3 4 5 6 ...
Page 164: ...Installation 162 ...
Page 165: ...Maintenance 5 ...
Page 166: ......
Page 184: ...Maintenance 182 1 2 7 9 8 10 3 4 5 6 ...
Page 188: ...Maintenance 186 Figure 5 8 Replacing the Torch 1 3 4 6 8 2 9 10 5 7 ...
Page 272: ...Maintenance 270 ...
Page 273: ...Troubleshooting 6 ...
Page 274: ......
Page 293: ...Troubleshooting 291 Figure 6 1 Normal Plasma Conditions at 0 torch position ...
Page 294: ...Troubleshooting 292 Figure 6 2 Normal Plasma at 3mm position ...
Page 297: ...Troubleshooting 295 Figure 6 5 Injector too far forward ...
Page 298: ...Troubleshooting 296 Figure 6 6 No Aux Gas Torch may be glowing ...
Page 299: ...Troubleshooting 297 Figure 6 7 Air Leak or Spray Chamber Temperature too high ...
Page 300: ...Troubleshooting 298 Figure 6 8 Thin Plasma due to leak in Plasma Gas Line ...
Page 308: ......