Installation
149
Adjusting the Sampling Probe Height
1. Install the rinsing-port rack on the autosampler.
See the procedure
Assembling Sample Racks
starting on page 61
i
n the
S10
Autosamplers User’s Guide (Part No. 09931133)
for more information.
2. Place the rinsing port in the rinse location in the sample rack.
See the procedure
Installing the Rinsing Port
on the Rack
starting on page 55 in
the
S10 Autosamplers User’s Guide (Part No. 09931133)
for more information.
3. In the ICP software, from the main menu make the following selections:
4. If the sample probe is suspended over the rinse position select the
Instrument
tab -> Load Tray
to lower the probe. If the sample probe is not at the rinse
position select the
Instrument tab -> Go To Wash
to move it there.
5. Carefully push the sampling probe down until it bottoms in the rinsing port. Then
pull the sampling probe up by about 15mm as shown in the following figure. Use
the O-ring on the probe to prevent it from slipping down. The sampling probe is
now in the correct height.
6. To remove the sample rack, first select
Instrument
tab
->Load Tray
to raise the
sampling probe, then remove the rack.
Note
The procedure described below assumes that you are using a spectrometer
operated via ICP software. If you are using an AA or ICP-MS instrument,
please refer to the information on operating an autosampler in the online
help for the respective instrument.
Summary of Contents for AVIO 200
Page 1: ...AVIO 200 SPECTROMETER Hardware Guide ICP OPTICAL EMISSION ...
Page 2: ......
Page 3: ...AvioTM 200 Spectrometer Customer Hardware and Service Guide ...
Page 12: ...Contents 10 ...
Page 30: ...28 ...
Page 31: ...Safety Practices 1 ...
Page 32: ...30 Safety Practices ...
Page 56: ...54 Safety Practices ...
Page 57: ...Preparing Your Laboratory 2 ...
Page 58: ...Preparing Your Laboratory 56 ...
Page 70: ...Preparing Your Laboratory 68 ...
Page 71: ...System Description 3 ...
Page 72: ......
Page 97: ...Installation 4 ...
Page 98: ......
Page 119: ...Installation 117 Figure 4 8 Replacing the torch 1 3 4 6 8 2 9 10 5 7 ...
Page 121: ...Installation 119 1 2 7 9 8 10 3 4 5 6 ...
Page 164: ...Installation 162 ...
Page 165: ...Maintenance 5 ...
Page 166: ......
Page 184: ...Maintenance 182 1 2 7 9 8 10 3 4 5 6 ...
Page 188: ...Maintenance 186 Figure 5 8 Replacing the Torch 1 3 4 6 8 2 9 10 5 7 ...
Page 272: ...Maintenance 270 ...
Page 273: ...Troubleshooting 6 ...
Page 274: ......
Page 293: ...Troubleshooting 291 Figure 6 1 Normal Plasma Conditions at 0 torch position ...
Page 294: ...Troubleshooting 292 Figure 6 2 Normal Plasma at 3mm position ...
Page 297: ...Troubleshooting 295 Figure 6 5 Injector too far forward ...
Page 298: ...Troubleshooting 296 Figure 6 6 No Aux Gas Torch may be glowing ...
Page 299: ...Troubleshooting 297 Figure 6 7 Air Leak or Spray Chamber Temperature too high ...
Page 300: ...Troubleshooting 298 Figure 6 8 Thin Plasma due to leak in Plasma Gas Line ...
Page 308: ......