![PerkinElmer AVIO 200 Hardware Manual Download Page 255](http://html1.mh-extra.com/html/perkinelmer/avio-200/avio-200_hardware-manual_1530192255.webp)
Maintenance
253
5. Tighten the adjustment screw
slowly
(turn clockwise) until droplets just start
forming at the end of the SAMPLE OUT side of the sample tubing.
6. Tighten the adjustment screw until you see a smooth flow of bubbles (3 turns on
the peristaltic pump). Add an additional three more turns.
7. Turn off the pump. The tension is now correctly set. No bubbles should be seen
in the sample tubing once the sample tube is kept in the solution.
Cleaning the Pump
Periodically inspect the peristaltic pump rollers to make sure they are clean and move
freely. Clean the exterior of the pump with a cloth moistened with water.
The pump head on the peristaltic pump can be removed and cleaned if necessary. A
replacement pump head is available (Part No. 09220266).
Removing the Pump Head on the Peristaltic Pump
To remove the pump head on the peristaltic pump:
1. Remove the pump tubing.
2. Unscrew the large screw in the center of the pump head. Place the screw and
washer in a secure place.
3. Lift the pump head off. To clean the pump head, use water or a mild solvent and
dry thoroughly.
4. Return the pump head to the pump. Orient the pump head by aligning the notch
on the bottom of the pump head with the pin on the shaft. The pump head must
click in this slot for a proper fit. Replace the washer and tighten the screw.
Caution
Attention
If a spill occurs on the pump, immediately clean it up.
A corrosive spill
could damage the pump.
Si un déversement se produit sur la pompe, nettoyer immédiatement
vers le haut.
Un déversement corrosif pourrait endommager la pompe.
Summary of Contents for AVIO 200
Page 1: ...AVIO 200 SPECTROMETER Hardware Guide ICP OPTICAL EMISSION ...
Page 2: ......
Page 3: ...AvioTM 200 Spectrometer Customer Hardware and Service Guide ...
Page 12: ...Contents 10 ...
Page 30: ...28 ...
Page 31: ...Safety Practices 1 ...
Page 32: ...30 Safety Practices ...
Page 56: ...54 Safety Practices ...
Page 57: ...Preparing Your Laboratory 2 ...
Page 58: ...Preparing Your Laboratory 56 ...
Page 70: ...Preparing Your Laboratory 68 ...
Page 71: ...System Description 3 ...
Page 72: ......
Page 97: ...Installation 4 ...
Page 98: ......
Page 119: ...Installation 117 Figure 4 8 Replacing the torch 1 3 4 6 8 2 9 10 5 7 ...
Page 121: ...Installation 119 1 2 7 9 8 10 3 4 5 6 ...
Page 164: ...Installation 162 ...
Page 165: ...Maintenance 5 ...
Page 166: ......
Page 184: ...Maintenance 182 1 2 7 9 8 10 3 4 5 6 ...
Page 188: ...Maintenance 186 Figure 5 8 Replacing the Torch 1 3 4 6 8 2 9 10 5 7 ...
Page 272: ...Maintenance 270 ...
Page 273: ...Troubleshooting 6 ...
Page 274: ......
Page 293: ...Troubleshooting 291 Figure 6 1 Normal Plasma Conditions at 0 torch position ...
Page 294: ...Troubleshooting 292 Figure 6 2 Normal Plasma at 3mm position ...
Page 297: ...Troubleshooting 295 Figure 6 5 Injector too far forward ...
Page 298: ...Troubleshooting 296 Figure 6 6 No Aux Gas Torch may be glowing ...
Page 299: ...Troubleshooting 297 Figure 6 7 Air Leak or Spray Chamber Temperature too high ...
Page 300: ...Troubleshooting 298 Figure 6 8 Thin Plasma due to leak in Plasma Gas Line ...
Page 308: ......