Maintenance
236
Cleaning the GemCone Nebulizer
With routine use it may become necessary to perform a thorough cleaning.
1. Remove the nebulizer from the end cap (loosen both thumbscrews).
2. Disconnect the argon gas fitting and the capillary tube connection from the
nebulizer.
3. Immerse the nebulizer in a solution of liquid laboratory cleaner (such as
"Microcleaning" solution, Cole-Parmer). Use the recommended concentration
of cleaning solution.
4. Apply ultra-sonication to the solution containing the nebulizer for about five
minutes.
5. Rinse the nebulizer thoroughly using deionized water.
6. Dry the gas passages of the conespray nebulizer by connecting the argon gas
fitting and turning on the argon flow to 1 L/min. Allow the argon flow to dry the
internal passages for several minutes.
GemCone Liquid Fitting Maintenance
It is possible to remove the liquid inlet fitting to the nebulizer. The clear plastic
fitting has a knurled head and can be unscrewed by hand. Before reconnecting the
fitting, make sure that about 1-2 mm of tubing protrudes out from the sealing side of
the fitting. Screw the fitting back into the nebulizer and snug to hand tightness.
Summary of Contents for AVIO 200
Page 1: ...AVIO 200 SPECTROMETER Hardware Guide ICP OPTICAL EMISSION ...
Page 2: ......
Page 3: ...AvioTM 200 Spectrometer Customer Hardware and Service Guide ...
Page 12: ...Contents 10 ...
Page 30: ...28 ...
Page 31: ...Safety Practices 1 ...
Page 32: ...30 Safety Practices ...
Page 56: ...54 Safety Practices ...
Page 57: ...Preparing Your Laboratory 2 ...
Page 58: ...Preparing Your Laboratory 56 ...
Page 70: ...Preparing Your Laboratory 68 ...
Page 71: ...System Description 3 ...
Page 72: ......
Page 97: ...Installation 4 ...
Page 98: ......
Page 119: ...Installation 117 Figure 4 8 Replacing the torch 1 3 4 6 8 2 9 10 5 7 ...
Page 121: ...Installation 119 1 2 7 9 8 10 3 4 5 6 ...
Page 164: ...Installation 162 ...
Page 165: ...Maintenance 5 ...
Page 166: ......
Page 184: ...Maintenance 182 1 2 7 9 8 10 3 4 5 6 ...
Page 188: ...Maintenance 186 Figure 5 8 Replacing the Torch 1 3 4 6 8 2 9 10 5 7 ...
Page 272: ...Maintenance 270 ...
Page 273: ...Troubleshooting 6 ...
Page 274: ......
Page 293: ...Troubleshooting 291 Figure 6 1 Normal Plasma Conditions at 0 torch position ...
Page 294: ...Troubleshooting 292 Figure 6 2 Normal Plasma at 3mm position ...
Page 297: ...Troubleshooting 295 Figure 6 5 Injector too far forward ...
Page 298: ...Troubleshooting 296 Figure 6 6 No Aux Gas Torch may be glowing ...
Page 299: ...Troubleshooting 297 Figure 6 7 Air Leak or Spray Chamber Temperature too high ...
Page 300: ...Troubleshooting 298 Figure 6 8 Thin Plasma due to leak in Plasma Gas Line ...
Page 308: ......