11. X-ray imaging (using C-arm; PK-8*1*B/8*2*B only)
●
Please note that the part of the bed
frame marked
①
will appear in the X-
ray.
●
X-ray TV imaging cannot be
performed using the C-arm when the
stretcher is in the inclined position.
●
Please note that the exeternal
dimensions of the C-arm vary according
to the manufacturer. Select an arm that
meets the specified conditions.
①
Raise the stretcher to highest position, and
raise the transfer board / collapsible side-rail
at the side on which the C-arm will be
inserted.
②
Insert the C-arm projector into the space
between the main frame and the base cover
(the shaded area in the illustration to the left).
③
Adjust the C-arm to the position at which
the X-ray will be taken.
<PK-8*1*B>
<PK-8021B/8121B>
①
①
(Unit:cm)
(Unit:cm)
■
The C-arm enables X-ray TV imaging to be performed in the shaded area in the following
illustration.
33
A: At least 41
B: At least 57
C: Up to 43
D: Up to 29
Permissible C-arm dimensions (unit:cm)
●
If the transfer board
/
collapsible side-
rail on the oppsite side of the C-arm
proves to be an obstacle during X-ray
imaging, lower the transfer board
/
collapsible side-rail
.