When performing the height adjustment or inclination angle adjustment, pay attention
not to hit the wall, beam, or power outlet on the wall. The stretcher or wall may be
damaged or deformed. Make sure that the stretcher does not hit the wall, beam, or power
outlet during the height adjustment or inclination angle adjustment.
●
When setting up the stretcher, make sure that there is at least15 cm clearance from the
wall.The main frame moved back and forth during inclination angle adjustment and
could hit the wall, causing damage or deformation of the stretcher.
Secure at least 15 cm clearance from the wall.
■
To avoid possible malfunctions, do not install the stretcher in the following
locations.
●
Locations expose to direct sunlight
●
Locations near heat sources such as a stove or heater
●
Locations which is heavily electrically charged such as near an MRI.
●
Locations that has a high oxygen level
●
Locations exposed to excessive amounts of water vapor or oil vapor
●
Locations that tend to be exposed to high temperatures, high humidity, low
temperatures or dry conditions
※
The optimum ambient conditions for the stretcher are as follows.
Temperature 10
℃
to 40
℃
/ Humidity 30%RH to 75%RH / Atmospheric pressure
700hPa to 1060hPa.
●
Locations with large amount of dust, smoke, salt, sulfur, or corrosive substances
●
Locations with poor ventilation
●
Locations exposed to vibrations or shocks (including during transportaion)
●
Locations where the floor is not strong enough
●
Locations where the floor is not level
18