12. X-ray imaging (using X-ray casette; PK-8*3*B only)
■
X-ray imaging using the cassette-based system can be performed within the range indicated
by the shaded area in the diagram below.
■
Imaging can be performed with the back-section base in either raised or horizontal position.
■
Both of back and foot section using cassette which can be perform X-Ray.
●
Imaging cannot be performed when
the stretcher is in an inclined
(Trendelenburg or reverse
Trendelenburg) position.
●
Size............. Rectangular cassette: Up to 38.8
×
46.4cm
Square cassette: Up to 38.8
×
38.8cm
●
Thickness.... Up to 1.5cm
●
Weight........ Up to 4.0kg
[Setting the cassette]
●
Confirm the imaging part and the cassette size,
and insert the cassette into between the bases,
and move the cassette to proper position.
※
The cassette can be set on the center of the
stretcher by suiting the cassette end to
△
marking.
[Setting the cassette when raising the back
section base]
●
Confirm the imaging part and the cassette size,
and set the cassette on the X-ray cassette tray.
And then make sure that the knob bolts loosen,
and move the X-ray cassette tray to proper
position. Finally, fix the X-ray cassette tray by
rotating the knob bolts.
●
Make sure that the X-ray cassette tray
is fixed by tightening the knob bolts.
●
The cassette cannot be fixed. Do not operate
the stretcher on inserting the casette. There is a
risk of malfunction.
34
Rectangular cassette
Square cassette
Knob bolt
X-ray cassette tray
(Unit:cm)