Chapter 8. Maintenance
Chapter 8. Maintenance
8.1 Calibration
In the
CALIB
menu, the user can calibrate and trim the analog output and inputs and check other meter functions.
This chapter also covers updating
Sentinel LNG
software over the RS232 interface.
Before performing calibration of the
Sentinel LNG
, be sure the following equipment is available:
•
Current meter capable of precisely measuring 4 to 20 mA current levels
•
PC with RS232 Cable and
Hyperterminal
software (available on Windows operating systems)
While following the programming instructions, refer to Figure 236 on page 217.
8.1.1
Updating Sentinel LNG Instrument Software
1.
To set up the RS232, connect one end of a 9-pin RS232 cable to the COM-1 serial port on a PC and the other end to
Com Port I/O on terminal block TB2 on the
Sentinel LNG
, as described in Chapter 2,
Installation
.
WARNING!
Use RS232 connections in Non-Hazardous areas only.
Note:
The RS232 cable connection is not limited to the COM-1 serial port of the PC. Connect the RS232 cable to any
available RS232 serial port on the PC, and follow the same instructions as for the COM-1 port.
2.
On your PC, set up the
Hyperterminal
program.
Note:
The use of
Hyperterminal
is shown here as an example. If you are using a different communications
software, see its manual for detailed instructions.
a.
From the PC Start Menu, click
Programs/Accessories/Communications/Hyperterminal
to open the Hyperterminal
window.
b.
If the call is not connected, click on
New Connection
and enter a name. Click OK.
c.
In the
Connect To
window, select COM 1 as the desired port.
d.
In the
Properties
window, set the following parameters:
• Bits per second: 115200
• Data bits: 8
• Parity: None.
• Stop Bits: 1
• Flow control: None
Summary of Contents for Sentinel LCT
Page 2: ......
Page 4: ...ii no content intended for this page...
Page 6: ...Preface iv Sentinel LNG User s Manual...
Page 12: ...Contents x Sentinel LNG User s Manual...
Page 26: ...Chapter 1 Features and Capabilities Sentinel LNG User s Manual 14...
Page 178: ...Chapter 5 Using PanaView 166 Sentinel LNG User s Manual...
Page 184: ...Chapter 6 MODBUS Communications 172 Sentinel LNG User s Manual...
Page 202: ...Chapter 7 HART Communications 190 Sentinel LNG User s Manual...
Page 220: ...Chapter 9 Troubleshooting 208 Sentinel LNG User s Manual...
Page 222: ...Appendix A Menu Maps 210 Sentinel LNG User s Manual...
Page 230: ...Appendix A Menu Maps Sentinel LNG User s Manual 218...
Page 232: ...Appendix B CE Mark Compliance and High Noise Areas 220 Sentinel LNG User s Manual...
Page 236: ...Appendix C Data Records 224 Sentinel LNG User s Manual...
Page 240: ...Appendix C Service Record 228 Sentinel LNG User s Manual...
Page 244: ...Index 232 Sentinel LNG User s Manual...
Page 246: ...Warranty 234 Sentinel LNG User s Manual no content intended for this page...
Page 247: ......