Chapter 3. General Programming
48
Sentinel LNG User’s Manual
3.7.2 Entering Signal Parameters
In the Signal
submenu, set parameters that affect the transducer signal:
CAUTION!
The SIGNAL default settings are suitable for most applications. Consult the factory before changing
either of these settings.
•
Error Allowed
•
Compensation Status (additional option)
While following the programming instructions, refer to Figure 231 on page 212.
Note:
From the
COMPOSITE
menu, proceed directly to Step 4. To select a different option, scroll to that option and
press
[ENT]
twice to select it (once to enter and again to confirm the selection).
To enter the Signal submenu:
1.
Press
[ESC]
. The
Sentinel LNG
enters the User Program.
2.
Press the [u] key until PROG is bracketed and press
[ENT].
3.
Use the [p] and [q] keys to scroll to Composite and press
[ENT].
4.
Scroll to the Signal option and press
[ENT].
5.
Select Error Allowed, press
[ENT]
, use the arrow keys to enter the new value, and press
[ENT]
to confirm the entry.
6.
If Dimensional Compensation is an included option, at the next parameter, Compensation Status, the screen
shows the two options, On or Off
.
Scroll to the desired option, and press
[ENT]
.
Entering parameters in the Signal
option has been completed. Press
[ESC]
to return to the
PROG
menu, or continue
pressing
[ESC]
until the display screen reappears.
3.7.3 Setting Up Input Data Feeds
In the Inputs
submenu, set the status for three types of inputs:
•
Temperature
•
Pressure
•
Density
1.
From the COMPOSITE menu, select Inputs, and press [ENT]. The next screen lists three types of inputs:
Temperature, Pressure, and Density.
2.
Select the input to be changed and press [ENT]. The input status is described by two options: Fixed or Active.
3.
To experience a stationary input, select Fixed, and press [ENT]. Press [ENT] again to reveal the Fixed input value.
4.
To set a different value, use the arrow keys to change the digits, and press [ENT].
5.
To experience an active input, select Active, and press [ENT]. Press [ESC] to return to the previous menu.
Summary of Contents for Sentinel LCT
Page 2: ......
Page 4: ...ii no content intended for this page...
Page 6: ...Preface iv Sentinel LNG User s Manual...
Page 12: ...Contents x Sentinel LNG User s Manual...
Page 26: ...Chapter 1 Features and Capabilities Sentinel LNG User s Manual 14...
Page 178: ...Chapter 5 Using PanaView 166 Sentinel LNG User s Manual...
Page 184: ...Chapter 6 MODBUS Communications 172 Sentinel LNG User s Manual...
Page 202: ...Chapter 7 HART Communications 190 Sentinel LNG User s Manual...
Page 220: ...Chapter 9 Troubleshooting 208 Sentinel LNG User s Manual...
Page 222: ...Appendix A Menu Maps 210 Sentinel LNG User s Manual...
Page 230: ...Appendix A Menu Maps Sentinel LNG User s Manual 218...
Page 232: ...Appendix B CE Mark Compliance and High Noise Areas 220 Sentinel LNG User s Manual...
Page 236: ...Appendix C Data Records 224 Sentinel LNG User s Manual...
Page 240: ...Appendix C Service Record 228 Sentinel LNG User s Manual...
Page 244: ...Index 232 Sentinel LNG User s Manual...
Page 246: ...Warranty 234 Sentinel LNG User s Manual no content intended for this page...
Page 247: ......