Chapter 5. Using PanaView
142
Sentinel LNG User’s Manual
5.9.3 Alarms Setup (cont.)
15.
From the
Alarm(FailSafe)
menu, change the
Trip Point
and/or
Dead Band
text box readings (if required) and click on
BACK
. Confirm the new entries by clicking
Yes
on the
Verification
screens.
16.
To set up the alarm direction, from the
Alarm Type Options
menu (see Figure 170 on page 138), click twice on
Direction...
The following screen appears.
Figure 177: Alarm Trip Direction Menu
17.
Click on
Units
and the following screen appears.
Figure 178: Alarm Trip Direction Channel Options
Summary of Contents for Sentinel LCT
Page 2: ......
Page 4: ...ii no content intended for this page...
Page 6: ...Preface iv Sentinel LNG User s Manual...
Page 12: ...Contents x Sentinel LNG User s Manual...
Page 26: ...Chapter 1 Features and Capabilities Sentinel LNG User s Manual 14...
Page 178: ...Chapter 5 Using PanaView 166 Sentinel LNG User s Manual...
Page 184: ...Chapter 6 MODBUS Communications 172 Sentinel LNG User s Manual...
Page 202: ...Chapter 7 HART Communications 190 Sentinel LNG User s Manual...
Page 220: ...Chapter 9 Troubleshooting 208 Sentinel LNG User s Manual...
Page 222: ...Appendix A Menu Maps 210 Sentinel LNG User s Manual...
Page 230: ...Appendix A Menu Maps Sentinel LNG User s Manual 218...
Page 232: ...Appendix B CE Mark Compliance and High Noise Areas 220 Sentinel LNG User s Manual...
Page 236: ...Appendix C Data Records 224 Sentinel LNG User s Manual...
Page 240: ...Appendix C Service Record 228 Sentinel LNG User s Manual...
Page 244: ...Index 232 Sentinel LNG User s Manual...
Page 246: ...Warranty 234 Sentinel LNG User s Manual no content intended for this page...
Page 247: ......