Chapter 3. General Programming
52
Sentinel LNG User’s Manual
c.
Drive Ctl-Select Open Drain or Push-Pull
3.9.2.1
Frequency
1.
Units - Choose the channel, parameter and units to output.
2.
Fbase - Enter the lowest value of the Frequency range you want to output.
3.
Fspan - Enter the highest value of the Frequency range you want to output.
4.
Value(min) - Enter the value of the parameter you want to correspond to the min. value of the frequency range.
5.
Value(max) - Enter the value of the parameter you want to correspond to the max. value of the frequency range.
6.
Choose the Error Level, the way the output will respond during an error condition. Choose Force LO, Force HI, Hold
Last, or Force Value. When the meter goes into Error, the choosen Error level will be output through the Frequency
Output.:
3.9.2.2
Totalizer
1.
Units - Choose the channel, parameter and units to output
2.
Unit/Pulse - Choose how many units of the chosen parameter are accumulated until a pulse is sent out.
3.
Pulse Width - Choose the length of each pulse output.
Note:
Make sure the meter is not going to be outputting more than one pulse during this time, as this could lead to
missed pulses.
4.
Polarity - Choose the Negative or Positive edge of the pulse.
Option
Output Response
Force Lo
Forces Output to Fbase.
Force Hi
Forces Output to Fspan.
Hold Last
Holds the last “good” reading.
Force Value
Enables the user to enter a value between the Frequency Range, to be output during an
error.
Summary of Contents for Sentinel LCT
Page 2: ......
Page 4: ...ii no content intended for this page...
Page 6: ...Preface iv Sentinel LNG User s Manual...
Page 12: ...Contents x Sentinel LNG User s Manual...
Page 26: ...Chapter 1 Features and Capabilities Sentinel LNG User s Manual 14...
Page 178: ...Chapter 5 Using PanaView 166 Sentinel LNG User s Manual...
Page 184: ...Chapter 6 MODBUS Communications 172 Sentinel LNG User s Manual...
Page 202: ...Chapter 7 HART Communications 190 Sentinel LNG User s Manual...
Page 220: ...Chapter 9 Troubleshooting 208 Sentinel LNG User s Manual...
Page 222: ...Appendix A Menu Maps 210 Sentinel LNG User s Manual...
Page 230: ...Appendix A Menu Maps Sentinel LNG User s Manual 218...
Page 232: ...Appendix B CE Mark Compliance and High Noise Areas 220 Sentinel LNG User s Manual...
Page 236: ...Appendix C Data Records 224 Sentinel LNG User s Manual...
Page 240: ...Appendix C Service Record 228 Sentinel LNG User s Manual...
Page 244: ...Index 232 Sentinel LNG User s Manual...
Page 246: ...Warranty 234 Sentinel LNG User s Manual no content intended for this page...
Page 247: ......