Chapter 9. Troubleshooting
200
Sentinel LNG User’s Manual
9.2.16 E20: Special Input Error
Problem:
This message indicates a special input error.
Cause:
The special input exceeds the specified limits for the analog inputs, or no input device is connected.
Action:
Check the input device and the connecting cable. Recalibrate the analog input.
9.2.17 E21: API Error
Problem:
This message indicates an error in the API calculations.
Cause:
The combination of pressure, temperature, density, and/or flow inputs causes the calculation to not
properly resolve.
Action:
The specific API error can be found in the API Info section of the meter's program.
9.2.18 E22: Degraded Performance Error
Problem:
The accuracy of the measurement is degraded.
Cause:
A loss of one inner and/or outer chord measurement results in the sister inner/outer chord's
measurement to be substituted, to provide a flow measurement with degraded accuracy.
Action:
Consult the factory.
9.2.19 E30: Channel Disabled
Problem:
The channel is not available.
Cause:
The channel has been turned off.
Action:
Enter the
PROGRAM
menu and enable the channel (see
Activating a Channel/Path (Status)
on page 43).
9.3 Displaying Diagnostic Parameters
The
Sentinel LNG
offers built-in Diagnostic Parameters to aid in the troubleshooting of transducer and/or electrical
problems. To access these parameters, do the following:
1.
From the display screen, press
[ESC].
Note:
If the display screen is locked, you will need to enter
[ESC], [ENT], [ESC]
and the security code. Refer to
Unlocking and Locking the
Sentinel LNG
on page 40 for details.
2.
Press the [
] key. The screen will highlight the channel you wish to check. Be sure the desired channel appears on
the screen (or change it, as discussed in Chapter 3).
3.
Press the [
] key to scroll to the measurement
entry in the upper right. Press
[ENT].
Be sure
Diagnostic
is
highlighted, and press
[ENT].
4.
Press the [
] key twice to access the units parameter. Press
[ENT].
Scroll to the desired
Diagnostic
unit (as
described in Table 34 on page 6) and press
[ENT].
5.
If desired, repeat steps
1
-
4
for the other channel.
Summary of Contents for Sentinel LCT
Page 2: ......
Page 4: ...ii no content intended for this page...
Page 6: ...Preface iv Sentinel LNG User s Manual...
Page 12: ...Contents x Sentinel LNG User s Manual...
Page 26: ...Chapter 1 Features and Capabilities Sentinel LNG User s Manual 14...
Page 178: ...Chapter 5 Using PanaView 166 Sentinel LNG User s Manual...
Page 184: ...Chapter 6 MODBUS Communications 172 Sentinel LNG User s Manual...
Page 202: ...Chapter 7 HART Communications 190 Sentinel LNG User s Manual...
Page 220: ...Chapter 9 Troubleshooting 208 Sentinel LNG User s Manual...
Page 222: ...Appendix A Menu Maps 210 Sentinel LNG User s Manual...
Page 230: ...Appendix A Menu Maps Sentinel LNG User s Manual 218...
Page 232: ...Appendix B CE Mark Compliance and High Noise Areas 220 Sentinel LNG User s Manual...
Page 236: ...Appendix C Data Records 224 Sentinel LNG User s Manual...
Page 240: ...Appendix C Service Record 228 Sentinel LNG User s Manual...
Page 244: ...Index 232 Sentinel LNG User s Manual...
Page 246: ...Warranty 234 Sentinel LNG User s Manual no content intended for this page...
Page 247: ......