Chapter 9. Troubleshooting
9.7 Uncertainty in Flow Rate for a Non-Insulated Meter
Dimensional compensation considers the effect on the geometrical dimensions of the meter due to the material
thermal expansion or contraction. The fluid temperature inside of the meter is measured and used for this purpose.
Under severe ambient conditions, eg. –40°C in winter, the fluid temperature can be very different, by up to 10°C, from
the actual wall temperature. This will cause about a 0.04% error in the flow rate measured. The Figure 229 below
indicates that the errors increase as the temperature difference between the fluid and the ambient for both insulated
and non-insulated meters. For non-insulated meters at large temperature differences, the error is about one order
higher than that for insulated meters. Therefore for applications in severe weather conditions, it is recommended to
insulate the meter to ensure the meter accuracy.
Figure 229: Example of Flow Error at Various Temperature Differences
-0.02
-0.01
0
0.01
0.02
0.03
0.04
0.05
-40
-20
0
20
40
60
80
100
120
Temperature Difference between Fluid and Ambinent (deg C)
E
rr
o
r in
F
lo
w
R
a
te
(%
)
SS306 with insulation
SS306 without insulation
Carbon Steel A106 with insulation
Carbon Steel A106 without insulation
Meter
Meter
meter
meter
Summary of Contents for Sentinel LCT
Page 2: ......
Page 4: ...ii no content intended for this page...
Page 6: ...Preface iv Sentinel LNG User s Manual...
Page 12: ...Contents x Sentinel LNG User s Manual...
Page 26: ...Chapter 1 Features and Capabilities Sentinel LNG User s Manual 14...
Page 178: ...Chapter 5 Using PanaView 166 Sentinel LNG User s Manual...
Page 184: ...Chapter 6 MODBUS Communications 172 Sentinel LNG User s Manual...
Page 202: ...Chapter 7 HART Communications 190 Sentinel LNG User s Manual...
Page 220: ...Chapter 9 Troubleshooting 208 Sentinel LNG User s Manual...
Page 222: ...Appendix A Menu Maps 210 Sentinel LNG User s Manual...
Page 230: ...Appendix A Menu Maps Sentinel LNG User s Manual 218...
Page 232: ...Appendix B CE Mark Compliance and High Noise Areas 220 Sentinel LNG User s Manual...
Page 236: ...Appendix C Data Records 224 Sentinel LNG User s Manual...
Page 240: ...Appendix C Service Record 228 Sentinel LNG User s Manual...
Page 244: ...Index 232 Sentinel LNG User s Manual...
Page 246: ...Warranty 234 Sentinel LNG User s Manual no content intended for this page...
Page 247: ......