Sentinel LNG User’s Manual
53
Chapter 3. General Programming
3.9.3 Alarms
1.
Select Status, and choose either OFF, Normal, or Failsafe.
a.
OFF - Disables alarm functionality.
b.
Normal - Selection for conventional operation.
c.
Failsafe - Selection for fail-safe operation.
2.
If Normal or Failsafe has been chosen, select Type.
a.
Level
Units - Defines the Unit of Measure for the alarm.
Trip Point - Defines the level where the alarm trips.
Dead Band - Defines the amount past the trip point where the alarm resets.
Trip Direction - Defines the direction of the alarm trip, above the set Trip Point or Below the set Trip Point
Error Level -
Reset - Reset would force the alarm state to reset on error or fault condition. Reset would be such that
NO would be NO, unless configured for failsafe.
Trip - Trip would force the alarm to Trip direction on error or fault condition.
Hold Last - Hold Last would hold the current state of the alarm on an error or fault condition.
b.
Direction
The alarm can be tripped on forward or reverse flow. It uses volumetric in m
3
/s. The alarm will look for a
positive or negative change in this flow reading.
c.
Fault Alarm
Trips the alarm on a particular fault, like E1 or E2. Can configure for individual faults, all faults or flow faults.
3.
Choose Latch State
a.
Latch - Enables the alarm to hold its state. This selection causes the User Intervene Indicator to be lit when an alarm
occurs.
b.
No Latch - The alarm is reset when the error condition clears.
c.
Clear Latch - Changes the latched state. This will turn off the User Intervene Indicator if the meter is no longer
in error.
Summary of Contents for Sentinel LCT
Page 2: ......
Page 4: ...ii no content intended for this page...
Page 6: ...Preface iv Sentinel LNG User s Manual...
Page 12: ...Contents x Sentinel LNG User s Manual...
Page 26: ...Chapter 1 Features and Capabilities Sentinel LNG User s Manual 14...
Page 178: ...Chapter 5 Using PanaView 166 Sentinel LNG User s Manual...
Page 184: ...Chapter 6 MODBUS Communications 172 Sentinel LNG User s Manual...
Page 202: ...Chapter 7 HART Communications 190 Sentinel LNG User s Manual...
Page 220: ...Chapter 9 Troubleshooting 208 Sentinel LNG User s Manual...
Page 222: ...Appendix A Menu Maps 210 Sentinel LNG User s Manual...
Page 230: ...Appendix A Menu Maps Sentinel LNG User s Manual 218...
Page 232: ...Appendix B CE Mark Compliance and High Noise Areas 220 Sentinel LNG User s Manual...
Page 236: ...Appendix C Data Records 224 Sentinel LNG User s Manual...
Page 240: ...Appendix C Service Record 228 Sentinel LNG User s Manual...
Page 244: ...Index 232 Sentinel LNG User s Manual...
Page 246: ...Warranty 234 Sentinel LNG User s Manual no content intended for this page...
Page 247: ......