Sentinel LNG User’s Manual
45
Chapter 3. General Programming
1.
Press
[ESC]
. The
Sentinel LNG
enters the User Program.
2.
Press the [u] key until PROG is bracketed and press
[ENT].
3.
Use the [p] and [q] keys to scroll to the desired Channel and press
[ENT].
4.
Scroll to the Signal option and press
[ENT].
5.
The prompt requests the Zero Cutoff. Near “zero” flow, the
Sentinel LNG
may have fluctuating readings due to
small offsets (caused by factors such as thermal drift in the fluid). The zero cutoff causes velocity measurements
less than the cutoff to be reported as zero. To set the cutoff, press
[ENT]
, and use the arrow keys to enter the new
value. Press
[ENT]
to confirm the entry.
Entering parameters in the Signal option has been completed. Press [ESC] to return to the PROG menu, or continue
pressing [ESC] until the display screen reappears.
Summary of Contents for Sentinel LCT
Page 2: ......
Page 4: ...ii no content intended for this page...
Page 6: ...Preface iv Sentinel LNG User s Manual...
Page 12: ...Contents x Sentinel LNG User s Manual...
Page 26: ...Chapter 1 Features and Capabilities Sentinel LNG User s Manual 14...
Page 178: ...Chapter 5 Using PanaView 166 Sentinel LNG User s Manual...
Page 184: ...Chapter 6 MODBUS Communications 172 Sentinel LNG User s Manual...
Page 202: ...Chapter 7 HART Communications 190 Sentinel LNG User s Manual...
Page 220: ...Chapter 9 Troubleshooting 208 Sentinel LNG User s Manual...
Page 222: ...Appendix A Menu Maps 210 Sentinel LNG User s Manual...
Page 230: ...Appendix A Menu Maps Sentinel LNG User s Manual 218...
Page 232: ...Appendix B CE Mark Compliance and High Noise Areas 220 Sentinel LNG User s Manual...
Page 236: ...Appendix C Data Records 224 Sentinel LNG User s Manual...
Page 240: ...Appendix C Service Record 228 Sentinel LNG User s Manual...
Page 244: ...Index 232 Sentinel LNG User s Manual...
Page 246: ...Warranty 234 Sentinel LNG User s Manual no content intended for this page...
Page 247: ......