L-718-8-E
CAUTION: The transmitter may be rotated over an 180° arc as shown in Fig. 4.10. It will not
turn beyond stop pins; do not force or the equipment will be damaged or fail.
(5) Confirming that the transmitter is secured in place, make conduit and wiring connections.
(6) Verify that the flowmeter operates properly.
4.7 How to Change Transmitter Orientation
To change transmitter orientation, follow the steps given below:
(1) Turn off power.
(2) Make preparations for conduit and cable so that they will by no means cause you any trouble while
changing transmitter orientation.
Work on a level plane with the transmitter up and the sensor secured so that loosening bolts will not
cause the transmitter to come off.
(3) Loosen bolts with hex wrench. See Fig. 4.10 (M8 hex socket head screws at four places).
(4) Rotate the transmitter to the desired direction and secure it with setscrews (four places). Make sure
that no clearance exists at this time as shown in Fig. 4.10.
Transmitter rotatable over an arc of 180°max.
Loosen four hex socket head screws (M8)
at the adapter and change the transmitter
orientation.
To Transmitter
To Sensor Unit
M8 setscrew
(4 places)
Fig.4.10 Changing Transmitter Orientation
WARNING
1. Do not attempt to work in a hazardous location.
2. Shock hazard! Be sure to turn off power before you start working.
WARNING
You may loosen bolts but never separate the transmitter from the sensor assembly.