91
21.Error List
EN
General Error Codes
If the errors cannot be solved by the suggested actions, contact OMRON.
Message Dialog
Action
E10
Main unit malfunction.
Turn off main power supply switch.
Turn off the power. Turn the power back on. If the
problem is not resolved, contact OMRON.
E19
End of Calendar.
Unable to continue.
Change the date and time settings in the
settings menu after power has been
restored to the system.
The calendar function cannot be used after
December 31, 2030.
E13
E20
Printer error.
Turn off main power supply switch.
Turn off the power. Turn the power back on. If the
problem is not resolved, contact OMRON.
E21
No printer paper.
Load paper.
Open the printer cover and load paper. (Refer to
Section 8.5.)
E22
Printer cover is open.
Close cover.
Close the printer cover.
E23
Printer jam.
Remove jammed paper.
Open the printer cover and remove the paper. (Refer
to Section 8.5.)
E42
This memory card has not been formatted.
Push the eject button, remove the memory
card and insert a new memory card.
Obtain a new memory card and insert it into the
memory card slot on the main unit.
E45
The measurement data has not been
saved.
Retake the measurement.
E46
This system was not properly terminated.
The latest measurement data has not been
saved.
Data was not saved since the power was
disconnected or interrupted during measurement.
E47
There is no memory card inserted in the
device.
Therefore the operation was terminated.
Memory card has not been inserted. Insert a
memory card into the slot on the right side of the
main unit.
E48
ID DATA has not been copied.
Try to copy the information again following the
procedure on Section 18.10. starting from the step
on copying ID information.
Replace clock battery.
Contact OMRON.
Insert memory card.
Memory card has not been inserted. Insert a
memory card into the slot on the right side of the
main unit.
This memory card cannot be used.
Format the memory card (Section 18.10.) or replace
it with a different memory card.
When the memory card is not properly ejected,
important DATA may be lost. Turn the power off and
back on in order to insert a memory card.
Turn off the power. Reinsert the memory card and
turn the power back on.
Memory capacity is insufficient.
Replace with empty memory card.
Eject the memory card and replace it with a new
one. (Section 18.10.)
Current memory card cannot be used.
Turn off the power. Reinsert the memory card and
turn the power back on.
This value is out of range.
The number just entered was not appropriate.
Reenter the number.
Summary of Contents for HEM-9000AI
Page 94: ...94 22 Specifications ...
Page 95: ...95 22 Specifications EN ...