84
19.Calibration Measurement
4.
Move the [
] cursor to [CHECK] under [SETTING
VALUE]. [Fig. 4]
5.
Press the SET button.
(The [CHECK] screen will be displayed.) [Fig. 5]
6.
Firmly wrap the cuff on a strong tube-like material.
7.
Close the exhaust valve of the rubber bulb, and inflate to an appropriate value (target
pressure). The pressure value is displayed.
8.
Compare values of the main unit and the mercury blood pressure monitor.
9.
Open the exhaust valve of the rubber bulb, and deflate.
10.
Press the F1 key to return to the default screen.
[Fig. 4]
[Fig. 5]
Check result
(1) In the measuring rule, the test tolerance of blood pressure monitor (accuracy of
pressure when shipped from the factory) should not exceed 3 mmHg.
(2) If the CHECK result is +/- 3 mmHg or above, contact the dealer where you purchased
this unit or OMRON.