89
21.Error List
EN
E66
Overpressure
Error.
Excessive pressure
was applied to the AI
pulse wave sensor
unit.
Measurement error Too
much pressure is on the
sensor.
• There is a possibility that the AI
pulse wave sensor hit against
bone or something hard. After
reviewing Section 9.2., retake
the measurement.
E67
Flash ROM
Aberration.
AI pulse wave
measurement unit
malfunction.
The AI pulse wave
measurement unit has
failed.
• Turn off the power. Turn the
power back on. If the problem is
not solved, contact OMRON.
E70
AI Pulse Wave
Sensor is
stored.
Measurement
started while the AI
pulse wave sensor
was still in the
storage clip.
Measurement began
while the AI pulse wave
sensor was stored.
• Follow the instructions for
measurement preparation.
Refer to Section 9.
E71
AI Pulse Wave
Sensor is out of
storage.
Put the AI pulse
wave sensor back in
the storage clip.
The AI pulse wave
sensor has been out of
storage for an extensive
period of time.
• While holding down the slide
stoppers on both sides, place
the AI pulse wave sensor in its
original position.
E72
No Pulse Wave
Found.
The pulse wave
cannot be accurately
detected.
The pulse wave is not
detectable.
• There is a possibility that the AI
pulse wave sensor is not on the
artery. Reattach the AI pulse
wave sensor unit properly.
Refer to Section 9.2. Retake the
measurement.
E73
Body
Movement
Error.
The pulse wave was
not accurately
detected due to body
movement or
vibration.
The pulse wave was not
accurately detected due
to body movement.
• Check for arrhythmia.
• Make sure that the patient does
not move their arm or body
during measurement.
E74
Operation
Aberration.
Abnormal operation
is detected. Hold
down pressure was
manually set outside
the range.
Measurement Error /
Detect abnormal
operation.
• After reviewing Section 14.,
retake the measurement.
E75
Too Much Time
for
Measurement.
Measurement was
interrupted due to
time-out failure.
Measurement time
exceeded 5 minutes.
Measurement Error /
Inflation time exceeds 5
minutes.
• Reattach the AI pulse wave
sensor unit properly. Refer to
Section 9.2. Retake the
measurement.
E76
Communication
Error.
Communication error
is detected between
the main unit and the
AI pulse wave
measurement unit.
The AI pulse wave
measurement unit is not
communicating with the
main unit.
• The main unit might not be
properly connected.
• After double-checking the cable
connection, retake the
measurement.
E77
Communication
Error.
The AI pulse wave
measurement unit was
not communicating with
the main unit.
• Retake the measurement.
E80
Best Pressure
Uncalculated.
The AI pulse wave
sensor unit could not
detect a pulse.
An irregular pulse wave
or similar problem
prevented the AI from
being measured.
• There is a possibility that the AI
pulse wave sensor is not on the
artery. After reviewing Section
9.2., retake the measurement.
Error
No.
Error Name
Error Message on
Display
Error Content
Action