4.Cleaning and Maintenance
EN
4. Cleaning and Maintenance
Maintenance prior to use
1. Check this device regularly.
2. When not in use for a long time, ensure that this device functions correctly and safely before
using it.
• Check all cord, cable and tube connections.
• Check that this device functions correctly by checking button functions.
• Check that the time displayed on the device is accurate.
3. Do not install or store this unit in the following conditions:
• In direct sunlight.
• In a location that is dusty or has high air salinity.
• On a sloping or vibrating object that creates impact
• Where chemicals are stored or combustible gas is generated.
• In high temperature or high humidity.
• Where water, liquid medications or chemicals may be spilled.
Maintenance after use
Store this device and the attachments after performing the proper maintenance and in an
orderly way.
Clean the surface of the main unit, the surface of the wrist rest, the outside of the AI pulse
wave measurement unit, the cuff and cables with the following procedure.
1. Use a damp cloth with 30 to 50% isopropyl alcohol or
70% ethyl alcohol for cleaning the surface.
DO NOT
use gasoline, thinner, benzene or highly
concentrated alcohol when cleaning this device.
Do not disinfect this unit by autoclave or gas
sterilization (EOG, formaldehyde, or high
concentration ozone), or do not implement ultraviolet
disinfection.
Do not wash or wet the wrist belt.
Do not wash or wet the cuff.
2. Wipe with a soft, dry cloth. [Fig. 1]
PLEASE NOTE
•Do not sterilize
A
R
T
.
INDEX
INDEX
M
AX
MAX
[Fig. 1]
Turn the power on.
Slide the pulse wave sensor into position as
illustrated in [Fig. 2].
Turn the power off.
Wipe the sensor gently with a dry clean cotton
swab or with a moistened cotton swab dipped in
diluted rubbing alcohol. [Fig. 2]
After cleaning is completed, turn the power on.
Slide the AI pulse wave sensor back in holder.
Note:
The sensor may be damaged or scratched if the
power is off when sliding pulse wave sensor.
[Fig. 2]
Be careful when cleaning the AI pulse wave sensor!