14.AI Measurement
Press the [
] buttons to display results from each
measurement.
Example: If 35 waveforms are measured:
34/35
35/35
AVG
1/35
2/35
The pulse wave number is displayed below the
waveform (in the centre lower-left part of the
screen). [Fig. 8]
8.
Press the START button to continue measuring. At
this time, the message “Start measurement with the
same ID?” will be displayed. Select “Yes” or “No”.
Select the START button [Yes] to begin
measurement with the current ID.
Select the STOP button [No] to erase the ID and
return to the default screen. Enter the ID for the next
patient.
Select the F1 key [CANCEL] to return to the
previous screen and change the measurement
mode. [Fig. 9]
[Fig. 8]
[Fig. 9]
Pulse wave number
If during measurement the waveform becomes
irregular or it is determined that the pressurization
value is incorrect, an error message will appear on
screen (E73 - 82). In this case, measure again.
Refer to
Section 21. A
n error message window
appearing in the middle of the screen can be
closed by pressing any button.