21.Error List
Error Codes for AI Pulse Wave Measurement
Error
No.
Error Name
Error Message on
Display
Error Content
Action
E50
AI Pulse Wave
Sensor Unit
Malfunction.
Abnormality is
detected in the
output of the AI
pulse wave sensor
unit.
An error has occurred in
the AI pulse wave
sensor unit.
• Turn the power off, turn it back
on and check the pulse wave
sensor unit. If this does not
solve the problem, contact
OMRON.
E51
AI Pulse Wave
Sensor Unit
Disconnected.
The signal from the
AI pulse wave
sensor unit cannot
be detected.
E52
AI Pulse Wave
Sensor
Malfunction.
Abnormality is
detected in the
output of the AI
pulse wave sensor
unit.
E53
AI Pulse Wave
Sensor Error.
The pulse wave
cannot be accurately
detected.
An error has occurred in
the AI pulse wave
sensor unit.
• There is a possibility the AI
pulse wave sensor is not on the
artery or that the wrist belt is
loose. Reattach the AI pulse
wave sensor unit properly.
Refer to Section 9.2. Retake the
measurement.
E54
AI Pulse Wave
Sensor Error.
Abnormality is
detected in the
output of the AI
pulse wave sensor
unit.
An error has occurred in
the AI pulse wave
sensor unit.
• Turn the power off and check
the pulse wave measurement
unit connection. If the error
cannot be solved by the
suggested action, contact
OMRON.
E55
AI Pulse Wave
Sensor Error.
E56
AI Pulse Wave
Sensor Unit
Malfunction.
There is a circuit failure
in the AI pulse wave
sensor unit.
• Turn off the main unit.
E60
Air Leakage
Error.
There is air leaking
in the AI pulse wave
measurement unit.
An error has occurred in
the AI pulse wave
measurement unit
pressure control
system.
• Turn off the power and check
the AI pulse wave
measurement unit air tube
connection.
E61
Deflation Error.
The cuff was not
deflated properly.
Pulse wave was not
measured.
E62
Rapid Exhaust
Error.
E63
Inflation Pump
Error.
The cuff was not
inflated properly.
Pulse wave was not
measured.
E64
Intake Pump
Error.
There is air leaking
in the AI pulse wave
measurement unit.
An error has occurred in
the AI pulse wave
measurement unit
pressure control
system.
• Check the AI pulse wave
measurement unit air tube
connection.
E65
AI Pulse Wave
Measurement
Unit Sensor
Error.
AI pulse wave
measurement unit
malfunction.
An error has occurred in
the AI pulse wave
measurement unit
pressure control
system.
• Turn off the power and check
the AI pulse wave
measurement unit air tube
connection.