81
18.Configuration Options
EN
4.
Delete all the data in the memory card.
1. Enter the Setup Screen. (Refer to Section 18.2.)
2. Move the [
] cursor to [OTHER SETTINGS] under the
menu column and press the SET or [
] button to move
the [
] cursor to [SETTING LIST].
3. Press the
[
]
key and move the [
] cursor to [FORMAT
MEMORY CARD]. [Fig. 5].
4. Press the SET button.
The confirmation message “All data will be deleted. Do
you wish to format?” will be displayed. Press the F2 key
to start formatting. [Fig. 6]
To stop formatting, press the F1 key [CANCEL].
Formatting the memory card will take approximately 3
seconds.
The Amount of Data That Can Be Stored on the Memory Card
If PULSE WAVE DATA is AVERAGE ONLY, data for approximately 6,400 measurements can be
saved on the enclosed memory card (capacity 128MB). If PULSE WAVE DATA is ALL, the data for
approximately 320 30 sec measurements can be saved.
5. Press the F1 key [CANCEL] to return to the default screen.
[Fig. 5]
[Fig. 6]
* Back up the data memory card on a computer regularly.
Refer to the instruction manual of your computer if you have
questions regarding how to save memory card data to your
computer.
* Do not use any memory cards except those recommended by OMRON.