6.2 Operating procedures performed directly on the
device
This section describes the operating procedures that can be performed directly on the
OTMC 100
device using the pushbutton. The pushbutton can be accessed after unscrewing the water-tight
membrane vent. Use a pointed tool such as a ball-pen to press the pushbutton.
NOTICE
Ingressing water may cause equipment damage.
► Do not forget to reinsert the membrane vent in order to restore watertightness.
► Tighten the membrane vent with a tightening torque of 0.6 to 0.8 Nm.
Membrane vent
detached
Pushbutton
LED
Pushbutton
6.2.1 Performing a reboot of OTMC 100
A device reboot can also be performed via the web interface (see
1. Press the pushbutton and release it immediately.
2. The LED goes off for approx. 1 s.
3. The LED lights up red during the boot process.
4. After approx. 15 s the LED changes to green to indicate operational readiness.
Operating OTMC 100
OMICRON
31