Insert the connecting pipe (4) to the clamping blocks
on
OTMC 100
and the mast clamp as shown in the
figure.
5
4
Align the
OTMC 100
device in an upright position as
shown in the figure and tighten both clamping screws
(5) of the clamping blocks evenly with a tightening
torque of 5 Nm.
Assemble the delivered waterproof RJ45 connector
to the Ethernet cable according to the accompanying
assembly instructions and insert it to the RJ45
connector on the base plate of
OTMC 100
. Use
suitable cable ties to fasten the Ethernet network
cable to the connecting pipe and the mast.
Always use the waterproof RJ45 Ethernet
connector supplied with
OTMC 100
in order
to maintain the watertightness of the
device. Do not use
OTMC 100
outdoors
without a waterproof connector inserted.
The clamping blocks provide holes for
horizontal and vertical mounting of the
connecting pipe. Therefore, it is also
possible to mount
OTMC 100
on the top of
a vertically mounted connecting pipe, as
shown in the figure on the right.
OTMC 100 Series User Manual
18