![background image](http://html1.mh-extra.com/html/omega/sbr-ew100/sbr-ew100_user-manual_4033456067.webp)
4-26
IMSBR08-E2
MC
Clears the message printout
buffer.
Syntax
M C p 1
<terminator>
p 1
Clear the message printout buffer (0)
Example
Clear the message printout buffer.
M C 0
VG
Resets the report data of the
periodic printout.
Syntax
V G p 1
<terminator>
p 1
Operation type
2
Reset the report data of the periodic
printout.
Example
Reset the report data of the periodic printout.
V G 2
Description This setting is valid when the recorder is set to print
the report data using periodic printout.
YC
Initializes settings.
Syntax
Y C p 1
<terminator>
p 1
Initialization type
0
Initialize the Setting mode and Basic
Setting mode settings.
1
Initialize the Setting mode settings.
Example
Initialize the Setting mode and Basic Setting mode
settings.
Y C 0
UY
Stops the record position
adjustment.
Syntax
U Y p 1
<terminator>
p 1
Fixed to 0.
Query
U Y ?
Example
Stop the record position adjustment.
U Y 0
Description Returns the execution status of the record position
adjustment in response to a query.
0: Stopped, 1: In progress
4.7 Output Commands
BO
Sets the byte output order.
Syntax
B O p 1
<terminator>
p 1
Byte order
0
Outputs the data MSB first.
1
Outputs the data LSB first.
Query
B O ?
Example
Output the data MSB first.
B O 0
Description This command applies to the byte order of numerical
data during BINARY output.
CS
Sets the checksum (/C3 option).
Syntax
C S p 1
<terminator>
p 1
Use/Not use the checksum
0
Not use
1
Use
Query
C S ?
Example
Use the checksum.
C S 1
Description This command can be used only on the RS-422A/485
communication interface.
IF
Sets the status filter.
Syntax
I F p 1
<terminator>
p 1
Status filter value
(0.0.0.0 to 255.255.255.255)
Query
I F ?
Example
Set the status filter value to 1.0.4.0.
I F 1 . 0 . 4 . 0
Description For details, see chapter 6.
Note
Initialization of BO/CS/IF Command Settings
•
For serial communications
Settings entered using the BO/CS/IF commands revert to
their initial values when the recorder is reset (when the
recorder is power cycled, or the user exits the basic setting
mode with the YE command).
•
Byte output order, checksum: 0
•
Status filter: 255.255.255.255
If the recorder is reset, you must restore these settings.
•
For Ethernet communications
Settings entered using the BO/IF commands revert to their
initial values when the connection to the recorder is cut.
After reconnecting the recorder, you must reenter the
settings.
CC
Disconnects an Ethernet
connection (/C7 option).
Syntax
C C p 1
<terminator>
p 1
Disconnect the connection (0)