Query
Y A ?
Example
Set the IP address to 192.168.111.24, the subnet mask
to 255.255.255.0, and the default gateway to 0.0.0.0.
YA 192.168.111.24,255.255.255.0,
0.0.0.0
Description The settings specified by this command and saved
using the XE command take effect after the recorder is
power cycled.
YN
Sets the DNS (/C7 option).
When not using the DNS
Syntax
Y N p 1
<terminator>
p 1
Use/Not use the DNS (OFF)
Query
Y N ?
Example
Do not use the DNS.
Y N O F F
When using the DNS
Syntax
Y N p 1 , p 2 , p 3 , p 4 , p 5
<terminator>
p 1
Use/Not use the DNS (ON)
p 2
Address of the primary DNS server
(0.0.0.0 to 255.255.255.255)
p 3
Address of the secondary DNS server
(0.0.0.0 to 255.255.255.255)
p 4
Domain suffix 1 (up to 64 characters)
p 5
Domain suffix 2 (up to 64 characters)
Query
Y N ?
Example
Use the DNS server at 192.168.0.1.
Y N O N , 1 9 2 . 1 6 8 . 0 . 1
Description The settings specified by this command and saved
using the XE command take effect after the recorder is
power cycled.
YD
Sets whether to use the login
function via communication
(/C7 option).
Syntax
Y D p 1
<terminator>
p 1
Use/Not use the login function via communication
(USE, NOT)
Query
Y D ?
Example
Use the login function via communication.
Y D U S E
Description The settings specified by this command and saved
using the XE command take effect after the recorder is
power cycled.
YQ
Sets the communication
timeout (/C7 option).
When not using the timeout
Syntax
Y Q p 1
<terminator>
p 1
Enable/Disable communication timeout (OFF)
Query
Y Q ?
Example
Disable timeout.
Y Q O F F
When using the timeout
Syntax
Y Q p 1 , p 2
<terminator>
p 1
Enable/Disable communication timeout (ON)
p 2
Timeout time (1 to 120) [minutes]
Query
Y Q ?
Example
Enable communication timeout and set the timeout
period to 3 min.
Y Q O N , 3
Description The settings specified by this command and saved
using the XE command take effect after the recorder is
power cycled.
YK
Sets keepalive (/C7 option).
Syntax
Y K p 1
<terminator>
p 1
Enable/Disable keepalive (ON, OFF)
Query
Y K ?
Example
Disable keepalive.
Y K O F F
Description The settings specified by this command and saved
using the XE command take effect after the recorder is
power cycled.
UA
Sets the record position.
Pen model
Syntax
U A p 1 , p 2 , p 3
<terminator>
p 1
Record position selection
Z E R O
0% position
F U L L
100% position
p 2
Pen number (1 to 4)
p 3
Record position adjustment value
When p1 =
Z E R O
SBR-EW100: An integer
between 0 and 70
SBR-EW180: An integer
between 0 and 180
When p1 =
F U L L
SBR-EW100: An integer
between –45 to 15
(reference value: 3026)
SBR-EW180: An integer
between –165 to 165
(reference value: 5447)
Example
Adjust the 0% position of pen 1. Set the record position
adjustment value to 20.
U A Z E R O , 1 , 2 0
Description • Check the recorded result and correct the record
position adjustment value.
• To end the adjustment, execute the UY0 command.
If you do not end the adjustment, you cannot change
the execution mode.
• The unit for p3 is 1/30 mm.