4-15
IMSBR08-E2
Commands
4
Example
Enable DST and set the DST start time to hour 0 on the
2nd Sunday of June and the DST end time to hour 0
on the 2nd Sunday of December.
T D U s e , J u n , 2 n d , S u n , 0 , D e c , 2 n d , S u n , 0
Description • The DST start time and end time cannot be set to
the same time.
• The TD command can be used on the SBR-EW100
and SBR-EW180 with system version 1.11 or later.
On the SBR-EW100 with system version 1.02 or
earlier, use the SS command.
SS
Sets the DST.
Syntax
S S p 1 , p 2 , p 3
<terminator>
p 1
Standard time/DST selection
W i n t e r
Standard time
S u m m e r
DST
p 2
Switch time setting ON/OFF (ON, OFF)
p 3
Date and time (fixed to the YY/MM/DD HH format)
Y Y
Year (00 to 99, the lower 2 digits)
M M
Month (01 to 12)
D D
Day (01 to 31)
H H
Hour (00 to 23)
Query
S S [ p 1 ] ?
Example
Switch to DST on 0 hour of June 1st, 2005.
S S S u m m e r , O N , 0 5 / 0 6 / 0 1 0 0
Description • Parameter p3 is not available when p2 is OFF.
• The SS command can be used on the SBR-EW100
with system version 1.02 or earlier.
SK
Sets the computation constant
(/M1 option).
Syntax
S K p 1 , p 2
<terminator>
p 1
Constant number (K01 to K30)
p 2
Constant (up to 11 characters)
The range is –9.9999E+29 to –1.0000E–30, 0,
1.0000E–30 to 9.9999E+29.
(The + sign of “E+” can be omitted.)
Query
S K [ p 1 ] ?
Example
Set 1.0000E
−
10 to constant K01.
S K K 0 1 , 1 . 0 0 0 0 E
−
1 0
Description This command cannot be specified while computation is
in progress.
SJ
Sets the timer used in TLOG
computation (/M1 option).
Syntax
S J p 1 , p 2 , p 3
<terminator>
p 1
Computation channel number
p 2
Timer number (Periodic, 1, 2)
p 3
Sum scale (for TLOG.SUM)
O F F
Integrates the data per scan interval.
/ s
Integrates the physical amount that are in
unit of /s.
/ m i n
Integrates the physical amount that are in
unit of /min.
/ h
Integrates the physical amount that are in
unit of /hour.
/ d a y
Integrates the physical amount that are in
unit of /day.
Query
S J [ p 1 ] ?
Example
Enable timer 1 on computation channel 0B. No sum
scale designation.
S J 0 B , 1 , O F F
Description • This command cannot be specified while
computation is in progress.
• About p3
The data for sum computation are sampled every
scan interval. For data with units such as /s, /min,
/h, and /day as in a flow rate, the data can be
summed over the unit time as shown below.
Parameter p3 is valid only for sum values.
OFF
Σ
(measured value)
/s
Σ
(measured value) X scan interval
/min
Σ
(measured value) X scan interval/60
/h
Σ
(measured value) X scan interval/3600
/day
Σ
(measured value) X scan interval/
(3600 X 24)
The unit of the scan interval is seconds.
CM
Sets the communication
input data (/M1 option).
Syntax
C M p 1 , p 2
<terminator>
p 1
Communication input data number
(see section 4.3)
p 2
Communication input data
The selectable range is –9.9999E+29 to –
1.0000E–30, 0, 1.0000E–30 to 9.9999E+29.
(The + sign of “E+” can be omitted.)
Query
C M ?
Example
Set 1.0000E–10 to communication input data number
C01.
C M C 0 1 , 1 . 0 0 0 0 E
−
1 0
FR
Sets the interval for acquiring
data to the FIFO buffer
Syntax
F R p 1
<terminator>
p 1
Pen model: 125ms, 250ms, 500ms, 1s, 2s,
2.5s, 5s, 10s
Dot model: 1s, 2s, 2.5s, 5s, 10s
Query
F R ?
Example
Set the acquiring interval to the FIFO buffer to 1 s.
F R 1 s
Description • Set the acquiring interval to an integer multiple of the
scan interval that is greater than the scan interval.
• The scan interval on the pen model is 125 ms.
• The scan interval on the dot model varies depends
on the integration time of the A/D converter as
follows:
When the integration time is 16.7 ms or 20 ms;
6 channels: 1 s
12, 18, or 24 channels: 2.5 s
4.4 Setting Commands