USG-400
Installation and Connection
3-4
ISSUE1
3 Connection of compatible electrosurgical
generator (when using the THUNDERBEAT)
WARNING
Never connect a cable other than the optional communication
cable to the ultrasonic generator. Otherwise, the ultrasonic
generator cannot perform appropriately and injury or equipment
damage may result.
CAUTION
Avoid applying excessive force on the communication cable.
Otherwise, wire disconnection or other failure may result.
Be sure to retain the docking plug cover. If it is lost, the
ultrasonic generator cannot be used in the standalone
configuration.
Be sure not to lose the tools such as screws and fixture
plates. If these items are lost, the ultrasonic generator cannot
be connected to the compatible electrosurgical generator.
Connect the docking connector and docking plug correctly.
Otherwise, malfunction or ultrasonic generator damage may result.
Avoid getting liquid inside the ultrasonic generator.
Otherwise, it may be damaged.
Avoid applying excessive force on the docking connector,
docking plug and fixture plates. Otherwise, they may be damaged.
Heed the following cautions when assembling the docking
connector and docking plug:
−
Be careful not to pinch a finger.
−
Be careful not to drop a screw in the area around the docking
plug.
Connect the docking plug to the docking connector properly.
Improper connection such as oblique insertion or contamination
may cause output irregularities.
Securely fix the fixture plates on the compatible electrosurgical
generator. Otherwise, the compatible electrosurgical generator may
fall.
Remove the fixture plate prior to lifting the compatible
electrosurgical generator. Otherwise, the fixture plates may be
damaged.
Do not remove screws other than those screws that are
instructed to do so in the manual.
Disconnect the docking connector and/or the docking plug
properly as described in “Disconnection” on page 3-10
Otherwise, the ultrasonic generator, docking connector and/or
docking plug may be damaged.
Before connecting the compatible electrosurgical generator,
be sure that the power cord of the ultrasonic generator is
unplugged from the AC power socket on the ultrasonic generator.
When removing the docking plug cover, be careful to maintain
pressure on the cover, because the docking plug is spring-loaded.
Summary of Contents for USG-400
Page 1: ...ISSUE 1 2011 11 MAINTENANCE MANUAL USG 400...
Page 2: ......
Page 6: ......
Page 10: ...USG 400 Product Outline 1 4 ISSUE1 4 System chart...
Page 11: ...USG 400 ISSUE1 2 1 Specifications Chapter 2 Specifications 1 Specifications 2 2...
Page 14: ......
Page 34: ......
Page 38: ......
Page 43: ...USG 400 ISSUE1 5 5 Inspection Figure 5 3 Figure 5 4...
Page 50: ...USG 400 Inspection 5 12 ISSUE1 Figure 5 12...
Page 59: ...USG 400 ISSUE1 6 1 Safety check Chapter 6 Safety check 1 Check procedures 6 2 2 Check Card 6 6...
Page 65: ...USG 400 ISSUE1 7 1 Block Diagram Chapter 7 Block Diagram 1 Block Diagram 7 2...
Page 89: ......