USG-400
Safety check
6-6
ISSUE1
2 Check Card
Date of check
Product
SN
Inspected by
Checked by
Inspection item
Criteria
Measurement
Equipment No.
result
Grounding Resistance The resistance must be 0.1
Ω
Ω
Normal Condition
500
μ
A
μ
A
Ground leakage
current
Single Fault
1000
μ
A
μ
A
Normal Condition
100
μ
A
μ
A
Enclosure leakage
current
Single Fault
500
μ
A
300
μ
A (UL Deviation)
μ
A
AC 10
μ
A
μ
A
Normal Condition
DC 10
μ
A
μ
A
AC 50
μ
A
μ
A
Patient leakage
current
Single Fault
DC 50
μ
A
μ
A