USG-400
ISSUE1 8-11
Troubleshooting
What to do when no error code is displayed
No error window is displayed and no alarm sounds in the following error status.
Perform the indicated remedial actions.
Irregularity Description
Possible cause
Remedial Actions
The power cord is not connected.
Connect the power cord.
The power cord is defective.
Replace the power cord.
The power is not switched ON.
Press the power switch to ON.
The ultrasonic generator
cannot be turned ON.
Ultrasonic generator malfunction.
Contact Olympus.
The transducer is not connected.
Connect the transducer.
Transducer malfunction.
The transducer is abnormal. Immediately
remove the THUNDERBEAT and/or
SONICBEAT from the body cavity and
replace the transducer.
THUNDERBEAT or SONICBEAT
instrument is not connected.
Connect the THUNDERBEAT or
SONICBEAT instrument.
THUNDERBEAT or SONICBEAT
instrument malfunction.
The THUNDERBEAT or SONICBEAT
instrument is defective. Remove the
THUNDERBEAT and/or SONICBEAT from
the body cavity and replace the
THUNDERBEAT or SONICBEAT
instrument.
Footswitch is not connected.
Connect the footswitch.
Footswitch malfunction.
Check the footswitch operation by referring
to Section 5.7, “Verification of the
high-frequency (RF bipolar) output (when
using the THUNDERBEAT)”. If it is
malfunctioning, replace it with a new one.
The power is not switched ON.
Press the power switch to ON.
Another instrument is generating
energy output.
It is not permitted to generate outputs
simultaneously from two instruments.
Generate output from one instrument at a
time.
Compatible electrosurgical
generator is generating the
output.
Stop using the compatible electrosurgical
generator (the THUNDERBEAT cannot be
used when the compatible electrosurgical
generator is in use).
The touch-screen of the
compatible electrosurgical
generator is being operated.
Stop operating the touch-screen of the
compatible electrosurgical generator (the
THUNDERBEAT cannot be used when the
touch-screen of the compatible
electrosurgical generator is being
operated).
Energy output cannot be
generated.
The other handswitch on the
THUNDERBEAT or SONICBEAT
or footswitch pedal than the
intended one is being pressed
(no activation).
Release all of the switches.
Summary of Contents for USG-400
Page 1: ...ISSUE 1 2011 11 MAINTENANCE MANUAL USG 400...
Page 2: ......
Page 6: ......
Page 10: ...USG 400 Product Outline 1 4 ISSUE1 4 System chart...
Page 11: ...USG 400 ISSUE1 2 1 Specifications Chapter 2 Specifications 1 Specifications 2 2...
Page 14: ......
Page 34: ......
Page 38: ......
Page 43: ...USG 400 ISSUE1 5 5 Inspection Figure 5 3 Figure 5 4...
Page 50: ...USG 400 Inspection 5 12 ISSUE1 Figure 5 12...
Page 59: ...USG 400 ISSUE1 6 1 Safety check Chapter 6 Safety check 1 Check procedures 6 2 2 Check Card 6 6...
Page 65: ...USG 400 ISSUE1 7 1 Block Diagram Chapter 7 Block Diagram 1 Block Diagram 7 2...
Page 89: ......