USG-400
Safety check
6-2
ISSUE1
1 Check procedures
1-1 Grounding
Resistance
confirmation
○
Criteria
The resistance between an accessible metal part of the USG-400 and the grounding
terminal of the meet must be 0.1
Ω
or less.
○
Inspection procedures
(1) Measure the resistance value between the inlet grounding terminal the exterior chassis
screw farthest from the inlet to apply 25A (non-loaded condition : 6A or less) for five to ten
seconds.
1-2
Ground leakage current confirmation
○
Criteria
The ground leakage current must not be over the value at the table 1-5.
○
Inspection procedures
(1) Make measurements at the grounding terminal of the inlet.
1-3
Enclosure leakage current confirmation
○
Criteria
The enclosure leakage current must not be over the value at the table 1-5.
○
Inspection procedures
(1) Make measurement at the exterior chassis screw farthest from the inlet.
Inlet grounding terminal
USG-400
Chassis screw
Chassis screw
Chassis screw
Summary of Contents for USG-400
Page 1: ...ISSUE 1 2011 11 MAINTENANCE MANUAL USG 400...
Page 2: ......
Page 6: ......
Page 10: ...USG 400 Product Outline 1 4 ISSUE1 4 System chart...
Page 11: ...USG 400 ISSUE1 2 1 Specifications Chapter 2 Specifications 1 Specifications 2 2...
Page 14: ......
Page 34: ......
Page 38: ......
Page 43: ...USG 400 ISSUE1 5 5 Inspection Figure 5 3 Figure 5 4...
Page 50: ...USG 400 Inspection 5 12 ISSUE1 Figure 5 12...
Page 59: ...USG 400 ISSUE1 6 1 Safety check Chapter 6 Safety check 1 Check procedures 6 2 2 Check Card 6 6...
Page 65: ...USG 400 ISSUE1 7 1 Block Diagram Chapter 7 Block Diagram 1 Block Diagram 7 2...
Page 89: ......