USG-400
ISSUE1 6-5
Safety
check
(4) Place the probe tip of SONICBEAT instrument in water.
(5) Turn on the power switch of USG-400, and then measure the leak current when the
SONICBEAT is activating at level 3( either handswitch can be used)
1-5
Leakage current criteria
Criteria
No Mode
Normal condition
Single fault condition
1
Ground leakage current
500
μ
A 1000
μ
A
2 Enclosure leakage current
100
μ
A
500
μ
A(others)
300
μ
A(UL Deviation)
3
Patient leakage current
10
μ
A 50
μ
A
USG-400
Leakage current measuring